FDA Adverse Event Malfunction Summary report: N

NITINOL TC ELECTRODE

MDR report key: 6642273 · Received June 14, 2017

Report

Report Number
3006630150-2017-02046
Event Type
Malfunction
Date Received
June 14, 2017
Date of Event
May 22, 2017
Report Date
May 22, 2017
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250013054
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED: MODEL #: TCN-10 LOT #: 021916, 013017, 121716 DESCRIPTION: NITINOL TC ELECTRODE, 100 MM.

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED: MODEL #: TCN-10 LOT #: 021916, 013017, 121716, 111516 AND 020416 DESCRIPTION: NITINOL TC ELECTRODE, 100 MM DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES FOUND THAT ALL DEVICES EXHIBITED DISCOLORED EPOXY, CRACKS AND CHIP OUTS. THE ROOT CAUSE OF THIS FAILURE WAS DUE TO THERMAL STRESS ASSOCIATED WITH AUTOCLAVE CYCLES. THE COLOR OF NEWLY CURED EPOXY IS AMBER/LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT SEVERAL ELECTRODES HAVE PROBLEMS WHERE THE EPOXY IS CHIPPED AROUND THE HUB WHERE FLUID CAN SEEP IN.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT SEVERAL ELECTRODES HAVE PROBLEMS WHERE THE EPOXY IS CHIPPED AROUND THE HUB WHERE FLUID CAN SEEP IN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422992 NITINOL TC ELECTRODE ELECTRODES GXI COSMAN MEDICAL TCN-10 020916 00813250013054

Patients

Seq Age Sex Outcome Treatment
1