SCREW HA5 - LENGTH 35
Report
- Report Number
- 3005180920-2017-00300
- Event Type
- Malfunction
- Date Received
- June 14, 2017
- Date of Event
- May 15, 2017
- Report Date
- June 14, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFO RECEIVED BY MAIL OF (B)(6) ON DATE THE 30TH MAY 2017: THERE IS NO WAY TO TELL WHICH SCREWS EXACTLY WERE USED AMONG THE FOLLOWING 5 LOTS OF THE SAME REFERENCE: (B)(4). BATCH REVIEWS PERFORMED ON 12 JUNE 2017. LOT 1416012: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14 APRIL 2015. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. LOT 1413504: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28 NOVEMBER 2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, THIS IS THE THIRD SIMILAR EVENT REPORTED ON ITEMS OF THE SAME LOT. LOT 1310157: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 29 MARCH 2013. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. LOT 106863: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25 OCTOBER 2010. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. LOT 1110551: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25 OCTOBER 2011. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. ON 14 JUNE 2017 THE R&D PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEMS AND COMMENTED AS FOLLOWS: ON THE CUTTING BLOCK INSPECTED DURING THE VISUAL INSPECTION, THERE ARE NO SIGNS OF DAMAGE OR PRESENCE OF METAL SHAVING. ON THE HEAD OF THE FIXATION SCREWS THERE ARE EVIDENCE SIGNS OF DAMAGE CAUSED DUE TO THE CONTACT BETWEEN THE SCREW AND THE CUTTING GUIDE DURING INSERTION. IT IS CLEAR THAT THE DEBRIS COMES FROM THE FIXATION SCREWS BECAUSE OF THE LESS HARDNESS MATERIAL (AISI630) COMPARED WITH THE CUTTING BLOCK ONE (AISI420B). TO SOLVE THIS PROBLEM, WE HAVE ALREADY INTRODUCED A NEW VERSION OF FIXATION SCREWS WITH REDUCED HEAD MADE IN AISI420FMOD AND WE CHANGED THE SHAPE OF THE HOLES ON THE CUTTING BLOCK INTRODUCING CHAMFER AND RADII IN ORDER TO REDUCE THE RISK OF CONTACT BETWEEN THE SCREW AND THE GUIDE DURING FIXATION.
AS THE 4-IN-1 CUTTING BLOCK WAS BEING SCREWED WITH THE HEX-DRIVER, A METAL SHAVING EMERGED AND FELL INTO THE WOUND. SHAVING WAS RETRIEVED FROM WOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 420493 | SCREW HA5 - LENGTH 35 | SURGICAL INSTRUMENT FOR KNEE | LXH | MEDACTA INTERNATIONAL SA | 1416012 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Other |