PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2017-65575
- Event Type
- Malfunction
- Date Received
- June 12, 2017
- Date of Event
- February 25, 2014
- Report Date
- February 25, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
BUTTON ERROR ALARM AND NO BUTTON RESPONSE DUE TO MOISTURE DAMAGE KEYPAD TRACE. SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS,BATTERY TUBE THREADS CRACKED,BROKEN BELT CLIP SLOT,CRACKED RESERVOIRTUBE LIP. JR (B)(6) 2014 THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A BUTTON ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 347 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413918 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |