VIDAS® TOXO IGM
Report
- Report Number
- 3002769706-2017-00133
- Event Type
- Malfunction
- Date Received
- June 9, 2017
- Report Date
- July 25, 2017
- Manufacturer
- BIOMERIEUX SA
- Product Code
- LGD
- PMA / PMN Number
- K141133
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A CUSTOMER IN FRANCE NOTIFIED BIOMÉRIEUX OF FALSE NEGATIVE RESULTS FOR EXTERNAL QUALITY CONTROLS (CTCB) ASSOCIATED WITH VIDAS® TOXO IGM (REFERENCE (B)(4)). AN INVESTIGATION WAS PERFORMED. THE REPORT FROM (B)(4) WAS REVIEWED, WHEREAS, CTCB TOXO 171 - SAMPLE 1712, WAS FOUND NEGATIVE BY A MAJORITY OF VIDAS® PARTICIPANTS WHEN IT WAS EXPECTED TO BE POSITIVE WITH VIDAS® TOXO IGM. RESULTS FOR CTCB TOXO 171 - SAMPLE 1712 WERE AS FOLLOWS : 2 PARTICIPANTS GAVE A POSITIVE RESULT AS EXPECTED, 18 PARTICIPANTS GAVE AN EQUIVOCAL RESULT, 81 PARTICIPANTS GAVE A NEGATIVE RESULT. A REVIEW OF QUALITY RECORDS CONFIRMED THE MANUFACTURING OF VIDAS® TOXO IGM LOTS 170608-0 / 171213-0 / 180213-0 AND 170718-0 WERE WITHIN SPECIFICATION. THE CTCB TOXO 171 SAMPLE 1712 WAS TESTED AND THE RESULT WAS POSITIVE ON VIDAS® TOXO IGM LOTS 170608-0, 171213-0 AND 180213-0. A COMPLEMENTARY AVIDITY TEST WAS PERFORMED FOR THE EXTERNAL QUALITY CONTROL CTCB TOXO 171 SAMPLE 1712, AND GAVE A HIGH AVIDITY RESULT. ACCORDING TO THE PACKAGE INSERT, A RESULT OF AVIDITY HIGHER THAN 0.300 INDICATES AN OLD INFECTION, OLDER THAN 4 MONTHS. THE CONCLUSION OF A PROBABLE OLD INFECTION IS COHERENT WITH THE AGE OF PREGNANCY INDICATED IN CTCB REPORT (14 WEEKS). HYPOTHEIS: EQA CTCB 171 SAMPLE 1712 CONTAINS RESIDUAL IGM WHICH ARE NOT DETECTED BY VIDAS® TECHNIQUE. VIDAS® TOXO IGM DETECTS IGM FOR A SHORTER PERIOD OF TIME THAN OTHER COMPETITORS. IT IS ACCEPTABLE TO OBTAIN A DOUBTFUL RESULT IN A CONTEXT OF OLD INFECTION, ESPECIALLY MORE SO WHEN THE SENSITIVITY IS HIGHER THAN 90% IN THE CONTEXT OF A RECENT INFECTION (CNR PUBLICATION - J. CLIN. MICROBIOL.-2016-VILLARD-3034-42) . THE EXTERNAL QUALITY CONTROL CTCB TOXO 171 SAMPLE 1712 IS COMPOSED OF HUMAN MATRIX. EQA ARE MANUFACTURED AND CANNOT BE CONSIDERED AS NATURAL SAMPLES FROM PATIENT. IN CONCLUSION, THERE IS NO PERFORMANCE ISSUE AND VIDAS® TOXO IGM PERFORMED AS EXPECTED.
A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF DISCREPANT RESULTS ASSOCIATED WITH VIDAS® TOXO IGM (REFERENCE (B)(4)). THE CUSTOMER REPORTED THAT NEGATIVE RESULTS WERE OBTAINED WHILE THE EXPECTED RESULT WAS POSITIVE AS PART OF EEQ CTCB 171. THERE IS NO PATIENT ASSOCIATED WITH THIS TEST RESULT BECAUSE THE TESTING WAS PART OF EEQ CTCB. AN INTERNAL BIOMÉRIEUX INVESTIGATION WILL BE INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413330 | VIDAS® TOXO IGM | VIDAS® TOXO IGM | LGD | BIOMERIEUX SA | 30202 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |