PRECISION SPECTRA
Report
- Report Number
- 3006630150-2017-02180
- Event Type
- Malfunction
- Date Received
- June 9, 2017
- Date of Event
- May 23, 2017
- Report Date
- May 23, 2017
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- UDI-DI
- 08714729767725
- PMA / PMN Number
- 030017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #: SC-2218-50 SERIAL #: (B)(4) DESCRIPTION: LINEAR ST LEAD, 50CM MODEL #: SC-1132 SERIAL #: 111548 DESCRIPTION: PRECISION SPECTRA IMPLANTABLE PULSE GENERATOR.
ADDITIONAL INFORMATION WAS RECEIVED THAT THE REASON FOR THE IPG REPLACEMENT WAS BECAUSE THE PATIENT WANTED A MAGNETIC RESONANCE IMAGING (MRI) COMPATIBLE SYSTEM. SC-2218-50 (B)(4) THE COMPLAINT WAS CONFIRMED. THREE CABLES WERE FRACTURED AT THE KINKED SECTIONS OF THE LEAD BODY WHERE A CLIK ANCHORED, BUT NO CABLES WERE EXPOSED AT THE SITE. SC-2218-50 (B)(4) VISUAL/X-RAY INSPECTIONS AND CONTINUITY TEST WERE PERFORMED TO ENSURE THE DEVICE INTEGRITY. NO ANOMALIES WERE FOUND. SC-1132 (B)(4) DEVICE EVALUATION INDICATED THAT THE DEVICE PASSED ALL TESTS PERFORMED.
A REPORT WAS RECEIVED THAT THE PATIENT WAS NOT GETTING SUFFICIENT COVERAGE DUE TO HIGH IMPEDANCE. PHYSICIAN SUSPECTED MALFUNCTION ON ONE OF THE LEADS. THE PATIENT UNDERWENT AN IPG AND LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.
A REPORT WAS RECEIVED THAT THE PATIENT WAS NOT GETTING SUFFICIENT COVERAGE DUE TO HIGH IMPEDANCE. PHYSICIAN SUSPECTED MALFUNCTION ON ONE OF THE LEADS. THE PATIENT UNDERWENT AN IPG AND LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413157 | PRECISION SPECTRA | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-2218-50 | NA | 08714729767725 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |