RENEW ENDOCUT SCISSORS TIP REUSABLE
MDR report key: 6629166
·
Received June 9, 2017
Report
- Report Number
- 1223422-2017-00066
- Event Type
- Injury
- Date Received
- June 9, 2017
- Date of Event
- February 20, 2017
- Report Date
- February 28, 2017
- Manufacturer
- MICROLINE SURGICAL, INC.
- Product Code
- GEI
- PMA / PMN Number
- K974066
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICES WERE RETURNED, DECONTAMINATED AND VISUALLY INVESTIGATED. UPON VISUAL INSPECTION THIS COMPLAINT HAS BEEN CONFIRMED. THE FOLLOWING RETURNED TIPS WERE RETURNED WITH A DEFECTIVE HEAT SHRINK. THE 00130041- 2 CONFIRMED, 00130689- 2 CONFIRMED, 00130280 - 4 CONFIRMED. MICROLINE INC., HAS NOTICED AN INCREASE OF REPORTED DEVICES WITH THIS SIMILAR FAILURE. A CORRECTIVE ACTION PLAN HAS BEEN ISSUED IN EFFORTS TO ADDRESS THE INCREASE OF THESE FAILURES. REF CAPA (B)(4).
Description of Event or Problem · 1
DURING A LAPAROSCOPIC CHOLECYSTECTOMY SURGICAL PROCEDURE, THE HEAT SHRINK FROM THE RENEW ENDOCUT SCISSOR TIP, AND FELL INTO A PATIENT. THERE WAS NO HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 411969 | RENEW ENDOCUT SCISSORS TIP REUSABLE | MANUAL DETACHABLE SURGICAL INSTRUMENTS | GEI | MICROLINE SURGICAL, INC. | 3141 | 00130041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |