HEARTWARE® VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2017-01824
- Event Type
- Injury
- Date Received
- June 8, 2017
- Date of Event
- May 13, 2017
- Report Date
- March 15, 2019
- Manufacturer
- HEARTWARE, INC
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF LOG FILE DATA. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. PER THE INSTRUCTIONS FOR USE (IFU): THROMBUS IS A POTENTIAL EVENT THAT MAY BE ASSOCIATED WITH USE OF THE PRODUCT. THE INSTRUCTIONS FOR USE (IFU) AND PATIENT MANUAL PROVIDE GUIDELINES ON PROPER USAGE OF THE HVAD SYSTEM AND PROGRAMMING HVAD PUMP PARAMETERS. MOREOVER, THE IFU PROVIDES INSTRUCTION TO FURTHER EDUCATE THE PATIENT ABOUT PRODUCT SAFETY, ALARM MANAGEMENT, AND ANTICOAGULATION RECOMMENDATIONS. THE INSTRUCTIONS FOR USE (IFU) ADDRESSES GUIDELINES FOR OPTIMAL PATIENT MANAGEMENT INCLUDING HAVING ADEQUATE AND STABLE PRELOAD AVAILABLE. HEARTWARE WILL SUBMIT A SUPPLEMENTAL REPORT WHEN NEW FACTS ARISES WHICH MATERIALLY ALTERS INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT.
THE PUMP HW4353 WAS NOT RETURNED FOR EVALUATION. REVIEW OF THE MANUFACTURING DOCUMENTATION CONFIRMED THAT THE ASSOCIATED DEVICE MET ALL REQUIREMENTS FOR RELEASE. REVIEW OF THE CONTROLLER LOG FILES REVEALED 32 HIGH WATT ALARMS AND AN INCREASE IN POWER CONSUMPTION WHICH CORRELATES WITH THE REPORTED "SUSPECTED THROMBUS" EVENT. OF NOTE, THE PATIENT WAS HOSPITALIZED AND WAS RECEIVING TPA TREATMENT. THERE IS NO EVIDENCE TO SUGGEST THAT A DEVICE MALFUNCTION CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE AVAILABLE INFORMATION, THE MOST LIKELY ROOT CAUSE OF THE HIGH POWER EVENT OBSERVED IN THE LOG FILES CAN BE ATTRIBUTED TO EXTERNAL FACTORS SUCH AS THROMBUS FORMATION/INGESTION. POSSIBLE CLINICAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT INCLUDE THE PATIENT¿S PRE-EXISTING HISTORY AND RELATED COMORBIDITIES, THE PROGRESSION OF THEIR UNDERLYING DISEASE, ISSUES RELATED TO THE THERAPEUTIC USE OF ANTICOAGULANT AND ANTIPLATELET MEDICATIONS AND THE PATIENT'S COMPLEX POST-OPERATIVE COURSE. THERE ARE POSSIBLE PATIENT, PHARMACOLOGICAL AND PROCEDURAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2017 WITH DARK URINE AND A SUSPECTED THROMBUS. THE PATIENT'S INTERNATIONAL NORMALIZED RATIO (INR) WAS SUPRATHERAPEUTIC ON ADMISSION. IT WAS NOTED THAT THE PATIENT MUST HAVE TAKEN ANTICOAGULATION WHEN THEY NOTICED THE DARK URINE AS THE SITE REPORTED NONCOMPLIANCE WITH ANTICOAGULATION. THE PATIENT REMAINS HOSPITALIZED TO ADMINISTER TISSUE PLASMINOGEN ACTIVATOR (TPA) AND LET THEIR INR DRIFT. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403331 | HEARTWARE® VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM | DSQ | HEARTWARE, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR | Hospitalization| L| R | DVBC3D4 ICD, 6947M62 LEAD |