FDA Adverse Event Malfunction Summary report: N

MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM

MDR report key: 6623192 · Received June 8, 2017

Report

Report Number
9610612-2017-00300
Event Type
Malfunction
Date Received
June 8, 2017
Date of Event
May 14, 2017
Report Date
August 14, 2017
Manufacturer
AESCULAP AG
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF AESCULAP AG (MANUFACTURER). EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.

Description of Event or Problem · 0

COUNTRY OF COMPLAINT: UNITED KINGDOM. IT WAS REPORTED THAT THERE WERE THREE DIFFERENT INCIDENTS WITH THREE DIFFERENT PATIENTS. INCIDENT DETAILS: (1) DURING A DENTAL SURGERY THE DRILL OVERHEATED AND IT BURNED THE INSIDE OF THE PATIENT'S MOUTH. THE DRILL STILL HAS PARTS OF THE SOFT TISSUE FROM THE PATIENT'S MOUTH. (2) DURING A DENTAL SURGERY THE HAND PIECE STARTED SMOKING, THE HAND PIECE WAS CHANGED AND THE SURGERY WAS ABLE TO BE COMPLETED. (3) DURING A DENTAL SURGERY THE OIL INSIDE THE DRILL STARTED LEAKING INSIDE THE PATIENT'S MOUTH. NO HARM TO THE PATIENT. ALL MED WATCH SUBMISSIONS RELATED TO THIS REPORT ARE: 9610612-2017-00296, 9610612-2017-00297, 9610612-2017-00298. COMPONENTS IN USE LISTED AS CONCOMITANT DEVICES ARE: GD450M / MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM (X3), GD678 / MICROSPEED UNI MICRO 150 MOTOR (X3), GD672 / MICROSPEED UNI MOTOR CABLE F/FOOT CTRL. (X3).

Additional Manufacturer Narrative · 1

INVESTIGATION: THE HANDPIECE WAS MANUFACTURED AND DELIVERED IN 2013, ACCORDING TO THE LASER MARKINGS. NO REPAIRS HAVE BEEN EXECUTED IN THE PAST. A FUNCTIONAL TEST WAS HARDLY POSSIBLE. THE BALL BEARINGS ARE RUNNING VERY ROUGHLY AND LOUDLY. ONE BALL BEARING HAS RUSTED-IN INSIDE THE HANDPIECE AND IS BLOCKED. THE SECOND BALL BEARING (TIP) IS BROKEN AND CORRODED. FURTHERMORE STAINING AND RUST CAN BE FOUND AT THE INTERIOR AND AT THE COLLET AS WELL. DAMAGES, CORROSION AND RUST CAN BE FOUND AT THE SURFACE OF THE TIP. CONCLUSION AND ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF OUR INVESTIGATION, THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY RELATED TO AN OVERLOAD USE AND TO AN INSUFFICIENT MAINTENANCE OF THE DEVICES. NO CAPA IS NECESSARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404711 MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM HIGH SPEED POWER SYSTEMS HWE AESCULAP AG GD450M

Patients

Seq Age Sex Outcome Treatment
1 Other GD450MM / MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM| GD672 / MICROSPEED UNI MOTOR CABLE F/FOOT CTRL| GD678 / MICROSPEED UNI MICRO 150 MOTOR