FDA Adverse Event Malfunction Summary report: N

MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM

MDR report key: 6619597 · Received June 7, 2017

Report

Report Number
9610612-2016-00324
Event Type
Malfunction
Date Received
June 7, 2017
Date of Event
October 7, 2016
Report Date
June 18, 2020
Manufacturer
AESCULAP AG
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

B1- MALFUNCTION ONLY, CONFIRMED. B5 - UPDATE. ACCORDING TO THE PROVIDED INFORMATION, THE GD450M HANDPIECE IS NOT RELATED TO THE MENTIONED OVERHEATING. IT SEEMS THAT THE USED DRILL BURR WAS THE REASON FOR THAT. THE DRILL BURR AND HANDPIECE WERE NOT RETURNED TO THE MANUFACTURER. AN INVESTIGATION WAS NOT PERFORMED. AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF AESCULAP AG (MANUFACTURER). EXEMPTION NUMBER: E2014018.

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF AESCULAP AG (MANUFACTURER). EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.

Description of Event or Problem · 0

CLARIFICATION WAS RECEIVED: THE OPERATING ROOM STAFF CONFIRMED THAT THERE WAS NO PATIENT HARM. THE DRILL BURR (UNSPECIFIED PRODUCT) OVERHEATED; THE BURR WAS EXCHANGED. SINCE THE BURR WAS SUSPECTED, THE HANDPIECE WAS NOT CONSIDERED TO BE A COMPLAINED DEVICE.

Description of Event or Problem · 0

COUNTRY OF COMPLAINT: (B)(6). IT WAS REPORTED THAT THE BURR OVERHEATED AND BURNED THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398956 MICRO-LINE STRAIGHT HDPC 1:1 F/2.35X70MM HIGH SPEED POWER SYSTEMS HWE AESCULAP AG GD450M

Patients

Seq Age Sex Outcome Treatment
1 Other