FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 661696 · Received January 9, 2006

Report

Report Number
2243072-2006-00013
Event Type
Other
Date Received
January 9, 2006
Date of Event
December 11, 2005
Report Date
January 6, 2006
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT HER METER GIVING INACCURATE READINGS - GETTING SOME LOW READINGS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READING (243) AS REFERENCE POINT VS BD READINGS (489, 128, 111) YIELDED SOME DATA POINTS IN THE C/D RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064328

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other