BI-METRIC XR-SERIES HIP SYSTEM FEMORAL STEM 131.5 NECK
Report
- Report Number
- 0001825034-2017-03682
- Event Type
- Injury
- Date Received
- June 6, 2017
- Report Date
- August 17, 2018
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- LPH
- PMA / PMN Number
- PK052089
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. (B)(4). COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS NOT CONFIRMED. 9MM STEM, WAS RETURNED AND EVALUATED AGAINST THE COMPLAINT. VISUAL INSPECTION FOUND THE STEM TO BE ASSEMBLED WITH THE RETURNED 28MM MODULAR HEAD. THE STEM IS IN OVERALL GOOD CONDITION. A SMALL AMOUNT OF DISCOLORATION, CONTACT DAMAGE, AND FOREIGN DEBRIS WERE FOUND WITHIN THE POROUS COATING. THE DEVICE WAS MEASURED ACCORDING TO THE CORRESPONDING INSPECTION. ALL MEASURED CRITERIA WERE FOUND TO BE WITHIN SPECIFICATION. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
(B)(4). CONCOMITANT PRODUCTS: X180410 279620 BI-METRIC XR-SERIES HIP SYSTEM FEMORAL STEM 131.5 NECK; X31-400009 132940 EXACT/ ALLIANCE BROACH PROVISIONAL 9MM X 125MM STAINLESS STEEL/ TITANIUM NITRIDE; X31-400010 578440 EXACT/ ALLIANCE BROACH PROVISIONAL 10X 130MM STAINLESS STEEL/ TITANIUM NITRIDE; 163662 561040 28MM MOD HD STD NECK TP1 TAPER. (B)(6). CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-03597, 0001825034-2017-03598.
IT WAS REPORTED THAT DURING A TOTAL HIP ARTHROPLASTY OPERATION THE SURGEON INSERTED A 9MM STEM IMPLANT INTO PATIENT'S MEDULLARY CAVITY. THE IMPLANT COULD NOT BE NEATLY FIT WITH THE CAVITY EVEN THOUGH THE SURGEON USED A RASP OF THE SAME IMPLANT SIZE. THE STEM WAS PROTRUDED BY A FEW MILLIMETERS FROM THE PROXIMAL FEMUR. THE SURGEON THEN TRIED THE INSERTION WITH 10MM STEM. AFTER THE SURGEON RASPED THE CAVITY WITH 10MM RASP HOWEVER, THE SAME PROBLEM OCCURRED AGAIN. SUBSEQUENTLY, ANOTHER 10MM STEM WAS USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 396721 | BI-METRIC XR-SERIES HIP SYSTEM FEMORAL STEM 131.5 NECK | PROSTHESIS, HIP | LPH | BIOMET ORTHOPEDICS | N/A | 265610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |