FDA Adverse Event Malfunction Summary report: N

CSK-TC ELECTRODE

MDR report key: 6614528 · Received June 6, 2017

Report

Report Number
3006630150-2017-02003
Event Type
Malfunction
Date Received
June 6, 2017
Date of Event
February 13, 2017
Report Date
February 13, 2017
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250010107
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MODEL #: CSK-TC10 LOT#: 080713 DESCRIPTION: CSK TC ELECTRODE, 10CM QUANTITY: 2 MODEL #:CSK-TC10 LOT#: 030613 DESCRIPTION: CSK TC ELECTRODE, 10CM QUANTITY: 1.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT MULTIPLE ELECTRODES WERE NO LONGER WORKING. DEVICE ANALYSIS REVEALED THAT ALL 3 ELECTRODES HAVE SHAFTS BROKEN FROM THE HUB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
395051 CSK-TC ELECTRODE ELECTRODES GXI COSMAN MEDICAL CSK-TC10 080713 00813250010107

Patients

Seq Age Sex Outcome Treatment
1