FDA Adverse Event Injury Summary report: N

G7 ACETABULAR SHELL

MDR report key: 6607874 · Received June 2, 2017

Report

Report Number
0001825034-2017-03505
Event Type
Injury
Date Received
June 2, 2017
Date of Event
May 2, 2017
Report Date
November 17, 2017
Manufacturer
BIOMET ORTHOPEDICS
Product Code
PBI
PMA / PMN Number
PK121874
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT HAS BEEN RETAINED BY THE PATIENT AND WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION; HOWEVER, AN INVESTIGATION HAS BEEN INITIATED. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP MDR WILL BE SUBMITTED. (B)(4). CONCOMITANT MEDICAL PRODUCTS - G7 ACETABULAR LINER P/N 010000848 L/N 3573746; G7 SCREW P/N 010000998 L/N 3624605; G7 SCREW P/N 010000999 L/N 3604098; FEMORAL STEM P/N 00771100600 L/N 61963053; BIOLOX FEMORAL HEAD P/N 00877403201 L/N 2811973. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-03545, 0001825034-2017-03506.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS UNABLE TO BE CONFIRMED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. REVIEW OF COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS NO TRENDS WERE IDENTIFIED. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE THREE MONTHS POST-IMPLANTATION DUE TO RANGE OF MOTION COMPLICATIONS, SQUEAKING DURING AMBULATION AND DISASSOCIATION OF THE LINER FROM THE ACETABULAR SHELL. THE LINER REPORTEDLY MIGRATED AROUND THE FEMORAL HEAD, RESULTING IN THE FEMORAL HEAD ARTICULATING WITH THE ACETABULAR METAL SHELL. THE SURGEON INDICATED THAT ONE OF THE ACETABULAR SCREWS MAY NOT HAVE BEEN FULLY INSERTED, WHICH WOULD INTERFERE WITH FULL SEATING OF THE LINER. DURING THE PROCEDURE, THE ACETABULAR SCREWS, LINER, APICAL HOLE PLUG AND CERAMIC HEAD WERE REMOVED AND REPLACED. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
389483 G7 ACETABULAR SHELL PROSTHESIS, HIP PBI BIOMET ORTHOPEDICS N/A 3626596

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R