Narratives
THE ONLY PLAUSIBLE CAUSE THAT THE NURSE AIDE (DEVICE OPERATOR) COULD COME UP WITH, IS THAT SHE HAD NOT PROPERLY PLACE THE LOOP OF THE SLING PAST THE SAFETY CLIP ON THE STAND. THE CONCLUSION REACHED BY THE NURSE AIDE IS THE ONLY EXPLAINABLE ANSWER TO HOW IT HAPPENED, AS BOTH THE LIFT AND THE SLING CHECKED OUT IN GOOD WORKING ORDER BY THE PEOPLE AT (B)(6) CARE CENTER, (B)(6) OF THE (B)(6) DEPARTMENT OF AGING AND DISABILITY'S, AND (B)(6) OF EZ WAY, INC.
ON 05/05/2017 (B)(6) OF OUR SERVICE DEPARTMENT RECEIVED A CALL FROM (B)(6), A NURSE SURVEYOR OF REGULATORY WITH THE (B)(6) DEPARTMENT OF AGING AND DISABILITIES SERVICES. (B)(6) WAS ASKING FOR THE PROOF OF A PART PURCHASE. (B)(6) TOLD (B)(6) THERE WAS AN ACCIDENT THAT HAD OCCURRED RESULTING IN A DEATH AT THE (B)(6) CARE CENTER AND WAS FOLLOWING UP TO SEE THAT THE FACILITY HAD BOUGHT NEW SAFETY CLIPS. (B)(6) INFORMED (B)(6) THAT THE FACILITY HAD MADE THEIR OWN YELLOW SAFETY CLIPS. AFTER RECEIVING THE CALL FROM (B)(6), (B)(6) CONTACTED (B)(6) AT (B)(6) CARE FACILITY TO FIND OUT MORE INFORMATION. IN AN EMAIL FROM (B)(6) 05/05/2017 AT 2:17 PM; (B)(6) DESCRIPTION IN HER OWN WORDS IS AS FOLLOWS: "NURSE AIDE HAD RESIDENT RAISED IN LIFT. RIGHT CANVAS HOOK CAME LOOSE FROM PIGTAIL HOOK. IN RE-ENACTMENT, NURSE AIDE DECIDED SHE MUST HAVE NOT HOOKED IT OVER THE CLIP, BUT RATHER ON TOP OF THE CLIP. THAT IS THE ONLY WAY IT COULD HAVE SLIPPED OFF WHILE IT HAD THE RESIDENT'S FULL WEIGHT. LIFT AND CANVAS HARNESS WERE FULLY EVALUATED AFTERWARD AND SHOW NO MECHANICAL OR WEAR/TEAR SIGNS. WHEN CANVAS HOOK CAME LOOSE, RESIDENT FELL AND HIT HEAD ON DOOR IMMEDIATELY BEHIND HIM. RESIDENT WAS PRONOUNCED A FEW MINUTES LATER."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 389329 | EZ WAY, INC. | EZ STAND PREMIER 400 LB | FSA | EZ WAY, INC. | 896 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Death |