FDA Adverse Event Malfunction Summary report: N

105" IV ADMINISTRATION SET

MDR report key: 6603046 · Received May 31, 2017

Report

Report Number
2245270-2017-00024
Event Type
Malfunction
Date Received
May 31, 2017
Date of Event
April 18, 2017
Report Date
June 26, 2017
Manufacturer
VYGON MFG
Product Code
FPA
PMA / PMN Number
K011336
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO VYGON FOR EVALUATION AS PART OF THE COMPLAINT INVESTIGATION. THE RESULTS OF THIS INVESTIGATION ARE STILL PENDING AND WILL BE COMMUNICATED TO FDA WITHIN THIRTY DAYS OF ITS CONCLUSION VIA FOLLOW-UP MDR.

Additional Manufacturer Narrative · 1

ONE BN-602, LOT# 1607093D WAS RETURNED FOR EVALUATION. THE PRODUCT WAS LEAK TESTED. THE COMPLAINT FOR FLUID LEAK WHERE THE TUBING MEETS THE DRIP CHAMBER WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD WAS CONDUCTED FOR THIS PRODUCT WITH THE LOT NUMBER PROVIDED. THE QC INSPECTION REPORT INDICATES ZERO FINDINGS FOR LEAK TESTING DURING THIS LOT MANUFACTURED IN AUGUST 2016. ALL PRODUCTION AND QC PERSONNEL ARE UP TO DATE WITH THEIR TRAINING. CORRECTIVE ACTION: AS AN IMMEDIATE ACTION VYGON HAS STOPPED MANUFACTURING WITH DRIP CHAMBERS FROM THIS SUPPLIER, AND CONVERTED ALL PRODUCTION BACK TO THE PREVIOUSLY QUALIFIED SUPPLIER. TRAINING WILL BE PERFORMED WITH ALL OPERATORS PERFORMING BONDING OPERATIONS. A CORRECTIVE ACTION WILL BE ISSUED INTO VYGON USA QUALITY MANAGEMENT SYSTEM TO DOCUMENT THE FORMAL TRAINING. VYGON USA WILL CONTINUE TO MONITOR THIS FAILURE FOR FUTURE OCCURRENCES. IN ADDITION, CAPA (B)(4) HAS BEEN OPENED TO FURTHER INVESTIGATE THE LEAKING DRIP CHAMBER ISSUE.

Description of Event or Problem · 1

DRIP SET UP & TUBING WAS LEAKING JUST BELOW THE DRIP CHAMBER WHERE THE TUBING MEETS THE DRIP CHAMBER.

Description of Event or Problem · 1

DRIP SET UP & TUBING WAS LEAKING JUST BELOW THE DRIP CHAMBER WHERE THE TUBING MEETS THE DRIP CHAMBER

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
383875 105" IV ADMINISTRATION SET INTRAVASCULAR ADMINISTRATION SET FPA VYGON MFG BN-602 1607093D

Patients

Seq Age Sex Outcome Treatment
1