FDA Adverse Event
Malfunction
Summary report: N
UP OVER BALKIN CONTRALATERAL CHECK FLO PERFORMER INTRODUCE
MDR report key: 660120
·
Received November 30, 2005
Report
- Report Number
- 660120
- Event Type
- Malfunction
- Date Received
- November 30, 2005
- Date of Event
- November 14, 2005
- Report Date
- November 28, 2005
- Manufacturer
- COOK INCORPORATED
- Product Code
- DYB
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MO, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
PHYSICIAN PERFORMING ANGIOPLASTY ON LOWER EXTREMITY, FOLLOWING CHECK ANGIOGRAM THE UP AND OVER SHEATH BROKE OFF THE HUB BECOMING DISLODGED IN THE PATIENT. AN ADDITIONAL PUNCTURE WAS MADE ON THE OPPOSING SIDE AND A SNARE WAS UTILIZED TO RETRIEVE AND REMOVE THE SHEATH. FOLLOW UP REVEALS THAT WITHIN 48 HOURS, THE PATIENT DEVELPOED A SMALL PSEUDOANEURYSM THOUGHT TO BE ASSOCIATED WITH THE BROKEN SHEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UP OVER BALKIN CONTRALATERAL CHECK FLO PERFORMER INTRODUCE | INTRODUCER SHEATH WITHOUT WIRE GUIDE | DYB | COOK INCORPORATED | RCFW-6.0-38-40-RB-BLKN | 1556876 OR 1457827 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |