FDA Adverse Event Injury Summary report: N

CARRIERE MOTION CLASS II

MDR report key: 6599788 · Received May 30, 2017

Report

Report Number
2081322-2017-00002
Event Type
Injury
Date Received
May 30, 2017
Date of Event
May 1, 2017
Report Date
May 1, 2017
Manufacturer
ORTHO ORGANIZERS, INC.
Product Code
EJF
UDI-DI
00190707023362
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ORTHODONTIST REMOVED AS MUCH COMPOSITE AS POSSIBLE AROUND UL3, UL6, UR4 AND UR6. ONCE THE EXCESS COMPOSITE WAS REMOVED, THE DENTAL ASSISTANT USED A BRACKET REMOVING PLIER AND SQUEEZED BETWEEN THE BONDING PAD AND TOOTH. A 1MM OF INCISAL ENAMEL UL3 WAS FRACTURED. A SMALL AMOUNT OF FLOWABLE COMPOSITE RESIN WAS BONDED TO THE ENAMEL FRACTURE. BECAUSE THE APPLIANCE WAS BONDED TO THE VERY INCISAL EDGE OF THE FACIAL SURFACE, THE ORTHODONTIST THOUGHT THE ENAMEL MAY HAVE BEEN TOO BRITTLE AT THE EDGE OF THE TOOTH. ADDITIONALLY, THE ORTHODONTIST SWITCHED THE BONDING AGENTS APPROXIMATELY 9 MONTHS AGO DUE TO EXCESSIVE BOND STRENGTHS HE WAS EXPERIENCING, WHICH MAY HAVE LED TO THE FRACTURED ENAMEL. DEVICE NOT RETURNED.

Description of Event or Problem · 1

TOOTH FRACTURE WHILE DEBONDING CARRIERE MOTION APPLIANCE WHEN REMOVING THE CARRIERE FROM #11 TOOTH, THE MESIAL EDGE OF THE TOOTH FRACTURED OFF WITH THE CARRIERE. THE PATIENT IS SYMPTOM FREE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379406 CARRIERE MOTION CLASS II BRACKET, METAL, ORTHODONTIC EJF ORTHO ORGANIZERS, INC. 424-925CN 00190707023362

Patients

Seq Age Sex Outcome Treatment
1 11 YR Required Intervention