C-LEG
Report
- Report Number
- 9615892-2017-00016
- Event Type
- Injury
- Date Received
- May 30, 2017
- Date of Event
- March 1, 2017
- Report Date
- May 30, 2017
- Manufacturer
- OTTO BOCK HEALTHCARE PRODUCTS GMBH
- Product Code
- ISY
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE EVALUATION OF THE KNEE JOINT SHOWED A DEFECT OF THE VALVE BLOCK OF THE HYDRAULIC UNIT (CORRESPONDING ERROR CODE EMERGED AT LEAST 17 TIMES). THIS FAILURE IS RELIABLE RECOGNIZED BY THE SYSTEM AND LEADS TO AN ACTIVATION OF THE SAFETY MODE. THE REPETITIVE WARNING SIGNALS AND SAFETY MODES OF THE DEVICE (WHICH POINT OUT THAT THE DEVICE SHOULD BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION) WERE IGNORED BY THE USER (87 SELFTEST FAILURES AND AT LEAST 17 POSITIONING FAILURES WITH SAFETY MODES ACTIVATED BETWEEN 12/2016 AND 04/2017 ACCORDING TO PRODUCT STATUS REPORT). ACCORDING TO IFU THE DEVICE SHOULD HAVE BEEN RETURNED EARLIER FOR INSPECTION PURPOSES. SEE INSTRUCTION FOR USE (USER) - C-LEG 3C98-3/3C88-3 (646D790-04-1503), ISSUED ON MAR. 05, 2015: SECTION 9.3: SAFETY MODE THE PRODUCT AUTOMATICALLY SWITCHES TO SAFETY MODE IF A CRITICAL SYSTEM FAULT OCCURS (E.G. FAILURE OF A SENSOR SIGNAL). SAFETY MODE REMAINS IN EFFECT UNTIL THE ERROR HAS BEEN RECTIFIED. DEFAULT DAMPING VALUES ARE ACTIVATED IN SAFETY MODE. THIS MAKES LIMITED WALKING POSSIBLE FOR THE USER EVEN THOUGH THE SYSTEM IS NOT ACTIVE. THE SWITCH TO SAFETY MODE IS INDICATED BY BEEPS AND VIBRATION SIGNALS IMMEDIATELY PRIOR TO SWITCHING (SEE PAGE 66). SAFETY MODE CAN BE DISABLED BY CONNECTING THEN DISCONNECTING THE BATTERY CHARGER. IF THE PRODUCT SWITCHES INTO SAFETY MODE AGAIN, THIS MEANS A PERMANENT ERROR EXISTS. THE PRODUCT MUST BE INSPECTED BY AN AUTHORISED OTTOBOCK SERVICE CENTRE.
INITIAL INFORMATION: KNEE IS VIBRATING AND BEEPING AND UNEXPECTEDLY LOCKING UP. WON'T HOLD A CHARGE, TRIED 3 DIFFERENT CHARGERS. FURTHER INFORMATION RECEIVED ON 05/29/2017: PT. WAS WALKING DOWN HOUSE STAIRS AND TALKING. PT. BRUISED FACE AND BROKE RIB.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 378621 | C-LEG | EXTERNAL ABOVE KNEE PROSTHESIS | ISY | OTTO BOCK HEALTHCARE PRODUCTS GMBH | 3C98-3=9.2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Other |