PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBE
Report
- Report Number
- 3012307300-2017-01131
- Event Type
- Death
- Date Received
- May 26, 2017
- Report Date
- January 2, 2018
- Manufacturer
- SMITHS MEDICAL ASD, INC.
- Product Code
- JOH
- UDI-DI
- 15021312006056
- PMA / PMN Number
- K944178
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL 510(K) NUMBER: K083641. CUSTOMER HAS NOT RETURNED THE DEVICE TO THE MANUFACTURER FOR DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE AND IS RETURNED AND EVALUATED, THE MANUFACTURER WILL FILE A FOLLOW-UP REPORT DETAILING THE RESULTS OF THE EVALUATION.
THREE DEVICES WERE RETURNED FOR EVALUATION. VISUAL INSPECTION FOUND THAT THE CUFF WAS CUT/TORN AT THE END OF THE DEVICE. IT WAS OBSERVED THAT THERE WERE TINY SCRATCHES NEXT TO THE RIPPED CUFFS. IT WAS SUSPECTED THAT THE DEVICES CAME INTO CONTACT WITH A SHARP OBJECT EXTERNALLY OR HARD CARTILAGE INTERNAL TO THE PATIENT. A REVIEW OF THE DEVICE HISTORY RECORD FOUND THAT ALL SPECIFICATIONS WERE WITHIN MANUFACTURING SPECIFICATIONS. DIMENSIONAL ANALYSIS OF THE CUFFS MET THE WALL THICKNESS SPECIFICATION. BASED ON THE EVIDENCE, A ROOT CAUSE WAS UNABLE TO BE DETERMINED.
SEE MFR: 3012307300-2017-02002, 3012307300-2017-01129, 3012307300-2017-01130, 3012307300-2017-01343, 3012307300-2017-01374, 3012307300-2017-01375, 3012307300-2017-01376, 3012307300-2017-01377, AND 3012307300-2017-01378 (SAME PATIENT).
IT WAS REPORTED THAT THE CUFF OF A PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBE "RUPTURED". NO INJURY WAS REPORTED. SEE MFR: 3012307300-2017-01129 AND 3012307300-2017-01130.
IT WAS ADDITIONALLY REPORTED THAT THE PATIENT PASSED AWAY UNEXPECTEDLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 374564 | PORTEX® BIVONA® ADULT TTS¿ TRACHEOSTOMY TUBE | TUBE TRACHEOSTOMY AND TUBE CUFF | JOH | SMITHS MEDICAL ASD, INC. | 3238414 | 15021312006056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |