FDA Adverse Event Malfunction Summary report: N

TEMPO

MDR report key: 6593343 · Received May 26, 2017

Report

Report Number
3007420694-2017-00128
Event Type
Malfunction
Date Received
May 26, 2017
Date of Event
August 1, 2016
Report Date
June 24, 2017
Manufacturer
ARJO MEB. AB LTD.
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PLEASE NOTE THAT PREVIOUS MEDWATCH REPORTS FOR THIS PRODUCT MAY HAVE BEEN SUBMITTED FOR THE MANUFACTURING SITE ARJO MED (B)(4) LTD (UNDER REGISTRATION #9617021). AS OF 06/15/2010, THAT NUMBER WAS DE-ACTIVATED DUE TO SITE NO LONGER BEING A MANUFACTURER AND UNTIL 2014 COMPLAINTS RELATED TO THESE PRODUCTS WERE HANDLED BY ARJO HOSPITAL EQUIPMENT (B)(4) AND ANY MEDWATCH REPORTS WERE SUBMITTED UNDER REGISTRATION #9611530 OR MEDIBO (MEDIBO MEDICAL PRODUCTS (B)(4) / 3004468271) AND AH (B)(4) (ARJOHUNTLEIGH (B)(4) INC. / 9681684). FROM 2014 AND GOING FORWARD COMPLAINTS RELATED TO THESE PRODUCTS ARE TO BE HANDLED BY ARJOHUNTEIGH (B)(4)'S COMPLAINT HANDLING ESTABLISHMENT AND ANY MEDWATCH REPORTS WILL BE SUBMITTED UNDER REGISTRATION #3007420694. ADDITIONAL INFORMATION WILL BE PROVIDED UPON CONCLUSIONS OF THE MANUFACTURER'S INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). ARJOHUNTLEIGH HAS BEEN INFORMED THAT DURING START OF RESIDENT TRANSFER FROM CHAIR WITH TEMPO LIFT AND NON ARJOHUNTLIEGH SLING, CLIPS OF SLING BECAME DETACHED FROM SPREADER BAR OF LIFT. ON NOTICING THIS, THE RESIDENT WAS IMMEDIATELY LOWERED BACK TO HIS CHAIR. NO CONSEQUENCES TO RESIDENT WERE REPORTED. AN INVESTIGATION WAS CARRIED OUT INTO THIS COMPLAINT. WHEN REVIEWING SIMILAR REPORTABLE EVENTS, WE HAVE FOUND A NUMBER OF CASES WITH SIMILAR FAULT DESCRIPTION (CLIP DETACHMENT). THE TREND OBSERVED FOR REPORTABLE COMPLAINTS WITH THIS FAILURE MODE IS CURRENTLY CONSIDERED TO BE LOW AND STABLE. NO MALFUNCTIONS REGARDING LIFT WERE REPORTED WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE EVENT. ACCORDING TO THE ABOVE THE LIFT WAS FOUND TO HAVE BEEN UP TO SPECIFICATION WHEN THE EVENT TOOK A PLACE. IT CAN BE ESTABLISHED THE LIFT AND SLING WERE BEING USED FOR PATIENT HANDLING AT THE TIME OF EVENT OCCURRENCE BUT IT APPEARS IT CONTRIBUTED TO THE EVENT DUE TO A USE ERROR. A SLING CLIP, ONCE CORRECTLY ATTACHED AND MONITORED TO STAY IN PLACE AS THE WEIGHT OF THE PERSON IN THE SLING IS GRADUALLY TAKEN UP, AS INDICATED TO BE REQUIRED IN THE LABELLING, IS LOCKED IN POSITION WITH THE WEIGHT OF THE PATIENT. IT IS NOT LIKELY TO COME OFF DURING ON LABEL USE. BASED ON PRODUCT KNOWLEDGE AND PREVIOUSLY MADE SIMULATIONS WE CAN STATE FOLLOWING: WHEN THE SLING CLIP IS NOT ATTACHED AND UNDER TENSION WITH THE WEIGHT OF THE PERSON IN THE SLING FROM THE START, A DROP CAN BE IMMEDIATE AFTER NOT BEING SUPPORTED BY THE CHAIR. IF THE LABELLING IS FOLLOWED THERE CAN BE NO ISSUE. HOWEVER, IT IS POSSIBLE FOR THE CAREGIVERS TO NOT HAVE FOLLOWED THE LABELLING AND NOT CHECKED THE CLIPS ARE CORRECTLY ATTACHED AND REMAIN IN TENSION AS THE WEIGHT OF THE RESIDENT IS GRADUALLY TAKEN UP. THE USER IS OBLIGED TO MONITOR THE CLIPS BECOMING UNDER TENSION WHEN THE WEIGHT OF THE PATIENT IS GRADUALLY BEING LOADED ON. THERE ARE ADDITIONAL SCENARIOS, THAT ALSO INVOLVE USE THAT IS NOT FOLLOWING THE IFU. DURING BEGINNING OF TRANSFER THE RESIDENT MUST BE TURNED IN THE CORRECT DIRECTION. AS A RESULT THE CAREGIVER MUST MANIPULATE THE SPREADER BAR THAT HOLDS THE SLING AND IS ABLE TO TURN FOR THIS PURPOSE. THE INTENDED AND LABELLED USE IS THAT THIS OCCURS BY OPERATING AND MANIPULATING THE SPREADER BAR ITSELF, AND NOT THE SLING NOR THE PERSON IN THE SLING. IF THIS LABELLING IS FOLLOWED THERE CAN BE NO ISSUE. HOWEVER, IT IS POSSIBLE FOR THE CAREGIVER TO NOT HAVE FOLLOWED THE LABELLING AND HAVE USED THE PERSON IN THE SLING TO MANIPULATE THE SPREADER BAR. IN THIS CASE THE CLIP COULD BE INADVERTENTLY PULLED OFF BY THE CAREGIVER WHILE USING THE SLING OR THE PERSON IN THE SLING FOR REPOSITIONING. ACCORDING TO THE INSTRUCTION FOR USE FOR TEMPO LIFT: "WARNING: IMPORTANT: ALWAYS CHECK THAT THE SLING ATTACHMENT CLIPS ARE FULLY IN POSITION BEFORE AND DURING THE COMMENCEMENT OF THE LIFTING CYCLE, AND IN TENSIONS AS THE PATIENTS WEIGHT IS GRADUALLY TAKEN UP." IT SHOULD BE EMPHASIZED THAT NON ARJOHUNTLEIGH SLING (GLOVE CLIP SLING) WAS INVOLVED IN THE EVENT. PLEASE NOTE THAT TEMPO INSTRUCTION FOR USE WARNS: "WARNING ONLY USE ARJO SUPPLIED SLINGS THAT ARE DESIGNED TO BE USED WITH TEMPO. THE SLING PROFILES ILLUSTRATED (..) WILL HELP TO IDENTIFY THE VARIOUS ARJO SLINGS AVAILABLE." THE LABELLING FOR THE LIFT DEVICE INDICATE THE SYSTEM SHOULD BE USED BY TRAINED PERSONNEL THAT ARE AWARE OF THE IFU CONTENTS. IN THIS CASE WE COME TO THE ONE CONCLUSION, NAMELY THAT THERE WAS A USE ERROR THAT CAUSED THE EVENT. FROM THIS EVALUATION IT WOULD APPEAR MOST LIKELY THAT THE EVENT WAS CAUSED BY THE USER NOT FOLLOWING THE IFU. NOTE THAT THE CUSTOMER WAS VISITED AND INTERVIEWED BY A LOCAL ARJOHUNTLEIGH REPRESENTATIVE. PROPER LIFTING PROCEDURE AS WELL AS CHOOSING SLING BEFORE TRANSFER IS TO BE COMMUNICATED TO THE CUSTOMER. WE FIND THIS COMPLAINT TO BE REPORTABLE TO THE COMPETENT AUTHORITIES.

Description of Event or Problem · 1

ON 28TH APRIL 2017, ARJOHUNTLEIGH WAS NOTIFIED THAT DURING THE START OF RESIDENT TRANSFER FROM CHAIR WITH ARJOHUNTLEIGH TEMPO LIFT AND NON ARJOHUNTLEIGH SLING (GLOVE CLIP SLING, CARE AND INDEPENDENCE SYSTEMS LTD), ALL FOUR SLING CLIPS BEGAN TO LIFT OFF THE ATTACHMENT ON THE BOOM OF THE HOIST (THE TOP TWO CLIPS TO A GREATER DEGREE THAN THE BOTTOM TWO). ON NOTICING THIS, THE SERVICE USER WAS IMMEDIATELY LOWERED BACK TO HIS CHAIR AND THE TRIAL WAS ABANDONED. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375862 TEMPO NON-AC-POWERED PATIENT LIFT FSA ARJO MEB. AB LTD.

Patients

Seq Age Sex Outcome Treatment
1 Other