ADVIA CENTAUR XP TOXOPLASMA G (TOXO G) ASSAY
Report
- Report Number
- 1219913-2017-00132
- Event Type
- Malfunction
- Date Received
- May 25, 2017
- Date of Event
- May 10, 2017
- Report Date
- June 19, 2017
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- LGD
- PMA / PMN Number
- K102681
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CAUSE FOR THE DISCORDANT TOXOPLASMA G (TOXO G) RESULTS IS UNKNOWN. SIEMENS HEALTHCARE DIAGNOSTICS IS INVESTIGATING. THE REAGENT LOT NUMBER USED FOR THE PATIENT TESTING HAS BEEN REQUESTED. THE IFU STATES IN THE LIMITATIONS SECTION: "IN GEOGRAPHIC REGIONS THAT HAVE AN APPARENT LOW PREVALENCE OF TOXOPLASMA IGG IN ASYMPTOMATIC POPULATIONS, THE POSITIVE PREDICTIVE VALUE OF ANY ASSAY IS REDUCED DUE TO THE INCREASED POSSIBILITY THAT A POSITIVE RESULT IS ACTUALLY FALSELY POSITIVE. AS WITH ALL IN VITRO DIAGNOSTIC ASSAYS, EACH LABORATORY SHOULD DETERMINE ITS OWN REFERENCE RANGE(S) FOR THE DIAGNOSTIC EVALUATION OF PATIENT RESULTS."
SIEMENS FILED THE INITIAL MDR 1219913-2017-00132 ON MAY 25, 2017. ON 06/07/2017 ADDITIONAL INFORMATION: PATIENT IS FEMALE, (B)(6). PATIENT IS PREGNANT, (B)(6). TREATMENT: CLEXANE. TOXOPLASMA G RESULTS (INDEX): ADVIA CENTAUR XP: 117.6 POSITIVE, ALTERNATE METHOD: NEGATIVE. THE QUALITY CONTROL WAS IN RANGE. THE REAGENT LOT NUMBER USED AT THE TIME OF TESTING WAS TOXO G LOT 220. EXPIRATION DATE: 11/10/2017. THE UDI # IS (B)(4). SAMPLE TYPE VACUTAINER TUBE WITH GEL BARRIER. THE PATIENT SAMPLE WAS NOT AVAILABLE FOR EVALUATION AT THE MANUFACTURER SITE. THE CUSTOMER TESTED 4015 NEGATIVE SAMPLES WITH LOT 061220, SO THEIR SPECIFICITY WOULD BE 99.95% (4015/4017). THE RELATIVE SPECIFICITY RESULTS FROM THE 3 STUDIES IN THE ADVIA CENTAUR XP TOXOPLASMA G INSTRUCTIONS FOR USE (IFU) (10629904 REVISION AA) RANGE FROM 99.3% - 99.8%. THE CAUSE OF THE ELEVATED RESULTS CANNOT BE DETERMINED. BASED ON THE AVAILABLE INFORMATION, THE ADVIA CENTAUR XP TOXOPLASMA IGG LOT 061220 IS MEETING THE SPECIFICITY CLAIMS IN THE IFU. THE CAUSE FOR THE DISCORDANT TOXOPLASMA G (TOXO G) RESULTS IS UNKNOWN. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
FALSE POSITIVE ADVIA CENTAUR XP TOXOPLASMA G (TOXO G) RESULTS WERE OBTAINED ON TWO SAMPLES FROM THE SAME PATIENT. THE PATIENT IS PREGNANT AND IS ON TREATMENT WITH HEPARIN. THE RESULTS WERE REPORTED TO THE PHYSICIAN AND WERE QUESTIONED. BOTH PATIENT SAMPLES WERE RUN ON TWO ALTERNATE METHODS AND THE RESULTS WERE NEGATIVE. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT TOXOPLASMA G RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 371088 | ADVIA CENTAUR XP TOXOPLASMA G (TOXO G) ASSAY | TOXOPLASMA GONDII IMMUNOASSAY | LGD | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |