FDA Adverse Event Injury Summary report: N

POWERED WHEELCHAIR SYSTEM

MDR report key: 6590485 · Received May 25, 2017

Report

Report Number
9615350-2017-00003
Event Type
Injury
Date Received
May 25, 2017
Date of Event
March 3, 2017
Report Date
May 25, 2017
Manufacturer
MOTION CONCEPTS LP
Product Code
ITI
PMA / PMN Number
K150574
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS SYSTEM WAS ORDERED BY THE DEALER (((B)(4).). THE CRITICAL CLIENT SPECS WERE REQUESTED AND PROVIDED FOR THE CLIENT WEIGHT, THE CLIENT FINISHED KNEE TO HEEL LENGTH, AND TOTAL HEIGHT. THE CLIENT'S BODY TYPE AND SPECIFICATION FOR SEAT SIZE, SEAT HEIGHT, WEIGHT AND DIMENSIONS WERE ALL WITHIN THE ORIGINAL WHEELCHAIR BASE MANUFACTURER'S PRODUCT OFFERING. FROM THE REPORT, WE SEE THAT THE END USER WAS INDEPENDENTLY TRANSFERRING TO THE OTHER CHAIR AND THE TRANSFER ACTIVITY WAS NOT DONE IN PRESENCE OF A QUALIFIED HEALTHCARE PROFESSIONAL. IT IS NOTED THAT THE MOTION CONCEPTS USER MANUAL PROVIDED WITH THIS SYSTEM INCLUDED THE FOLLOWING WARNING: "DETERMINE AND ESTABLISH YOUR PERSONAL SAFETY LIMITS BY PRACTICING BENDING, REACHING AND TRANSFERRING ACTIVITIES IN THE PRESENCE OF A QUALIFIED HEALTHCARE PROFESSIONAL BEFORE ATTEMPTING ACTIVE USE OF THE WHEELCHAIR". HOWEVER, IN THIS CASE IT WAS NOT FOLLOWED BY THE END USER. IT IS CLEAR, THAT THE INCIDENT WAS NOT AS RESULT OF A MALFUNCTION OF THE WHEELCHAIR.

Description of Event or Problem · 1

ON 27-APRIL 2017, MOTION CONCEPTS WAS NOTIFIED BY INVACARE CORPORATION OF AN ALLEGED INCIDENT INVOLVING ONE OF OUR TRXULMXT50(CG) SYSTEM. INVACARE WAS INFORMED BY THEIR DEALER, NATIONAL SEATING AND MOBILITY, THAT THE END USER WAS USING HER ROVI WHEN IT JUST STOPPED WORKING. THEN SHE DECIDED TO INDEPENDENTLY TRANSFER INTO HER BACK UP CHAIR. DURING HER TRANSFER, NO CONTACT WAS MADE WITH EITHER WHEELCHAIR AND SHE FELL. HER CAREGIVER, RACHEL, A FRIEND, AND END USER'S MOM WERE PRESENT DURING THE INCIDENT. THE END USER EMPHASIZED THAT THE INJURY WAS HER FAULT. SHE DOES NOT BLAME THE CHAIR FOR IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372880 POWERED WHEELCHAIR SYSTEM WHEELCHAIR ITI MOTION CONCEPTS LP TRX-ULMXT50(CG)

Patients

Seq Age Sex Outcome Treatment
1 63 YR Hospitalization