FDA Adverse Event Malfunction Summary report: N

TRIAGE TOX DRUG SCREEN PANEL

MDR report key: 6585214 · Received May 23, 2017

Report

Report Number
2027969-2017-00090
Event Type
Malfunction
Date Received
May 23, 2017
Report Date
April 24, 2017
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
DIO
PMA / PMN Number
K060791
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION CONCLUSION: THE CUSTOMER'S COMPLAINT OF FALSE NEGATIVE COC WAS NOT REPLICATED WITH IN-HOUSE TESTING OF RETAINS OF LOT W62468RB. MANUFACTURING BATCH RECORDS FOR LOT W62468RB WERE REVIEWED AND FOUND THAT THE LOT MET RELEASE SPECIFICATIONS. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED. THIS EVENT IS BEING REPORTED AS PART NUMBER 94400EU IS SAME/SIMILAR TO K060791.

Description of Event or Problem · 1

THE CUSTOMER REPORTED CONTINUALLY RECEIVING NEGATIVE COCAINE RESULTS USING THE TRIAGE TOX DRUG SCREEN PANEL. THE CUSTOMER PERFORMED CONFIRMATORY TESTING ON ONE PATIENT SAMPLE AFTER RECEIVING A NEGATIVE COCAINE RESULT USING THE TRIAGE TOX DRUG SCREEN PANEL. THE CUSTOMER RECEIVED A POSITIVE COCAINE RESULT OF 389 NG/ML USING THE SIEMENS DIMENSION EXL 200 ANALYZER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367261 TRIAGE TOX DRUG SCREEN PANEL ENZYME IMMUNOASSAY, COCAINE AND COCAINE METABOLITES DIO ALERE SAN DIEGO, INC. 94400EU W62468RB

Patients

Seq Age Sex Outcome Treatment
1 TRIAGE METERPRO, PN 55071, SERIAL # (B)(4)