FDA Adverse Event Malfunction Summary report: N

SYNGO IMAGING

MDR report key: 6584601 · Received May 23, 2017

Report

Report Number
3002808157-2017-78797
Event Type
Malfunction
Date Received
May 23, 2017
Date of Event
April 25, 2017
Report Date
April 27, 2017
Manufacturer
SIEMENS HEALTHCARE GMBH
Product Code
LLZ
PMA / PMN Number
K082430
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE ARE NO INJURIES ATTRIBUTED TO THE REPORTED EVENT. HOWEVER, CLINICALLY RELEVANT DATA LOSS OCCURRED . THIS REPORT WAS SUBMITTED MAY 23, 2017. CUSTOMER'S ADDRESS: (B)(6).

Description of Event or Problem · 1

A DATA LOSS IN THE SYNGO IMAGING APPLICATION OCCURRED FOR 62 OUT OF 601 IMAGES IN ONE SERIES ON (B)(6) 2017. NO DATA MIX-UP OCCURRED; NO PATIENT RESCAN WAS REQUIRED. DUE TO CHANGES ON THE POWER SUPPLY OF THE SYNGO IMAGING ENVIRONMENT PERFORMED BY THE CUSTOMER (B)(6) 2017, AN IMPROPER SHUTDOWN OF THE STORAGE TERM RAID AND THE SYNGO IMAGING APPLICATION OCCURRED. THIS RESULTED IN THE AFFECTED IMAGES BEING CREATED WITH 0 KB FILE SIZE IN THE SYNGO IMAGING'S SHORT TERM STORAGE PRIOR TO THEM BEING ARCHIVED IN THE LONG TERM ARCHIVE. THE AFFECTED SERIES WERE AVAILABLE IN THE SYNGO IMAGING VIEWER, HOWEVER, ONLY 539 IMAGES CAN BE DISPLAYED AND 62 IMAGES WERE CORRUPT. THESE CORRUPT IMAGES WERE FOUND DUE TO MAINTENANCE TASKS ON THE SYSTEM (B)(6) 2017. THE REPORTED INCIDENT OCCURRED IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
364902 SYNGO IMAGING PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM. LLZ SIEMENS HEALTHCARE GMBH 10014063

Patients

Seq Age Sex Outcome Treatment
1 Other