FDA Adverse Event Injury Summary report: N

VERSAFITCUP ACETABULAR SHELL Ø 46

MDR report key: 6583839 · Received May 23, 2017

Report

Report Number
3005180920-2017-00501
Event Type
Injury
Date Received
May 23, 2017
Date of Event
April 25, 2017
Report Date
May 23, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
PMA / PMN Number
K083116
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 05 MAY 2017 THE (B)(4) PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: LATE INFECTION IN THA, 1 YEAR AFTER PRIMARY OPERATION. INFECTION IS A KNOWN POSSIBLE ADVERSE EVENT FOLLOWING EVERY SURGERY, INCLUDING THA'S. TO DATE, THERE IS NO REASON TO SUSPECT THAT THE CAUSE MAY BE LINKED TO THE IMPLANTED DEVICES. BATCH REVIEWS PERFORMED ON 17 MAY 2017. LOT 153605: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21 OCTOBER 2015. EXPIRATION DATE: 2020-10-07. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. COCR BALL HEAD 12/14 Ø 22 SIZE S -2,5, CODE 01.25.124, LOT. 141035 (K080885) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30 APRIL 2014. EXPIRATION DATE: 2019-03-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VERSAFITCUP DOUBLE MOBILITY HC LINER Ø 46/22.2, CODE 01.26.2246MHC, LOT. 152686 (K092265) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22 OCTOBER 2015. EXPIRATION DATE: 2020-10-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. AMISTEM C, CEMENTED STEM SIZE 0 STD, CODE 01.18.150, LOT. 150877 (K103189) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13 MAY 2015. EXPIRATION DATE: 2020-03-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE INFECTION IS CONFIRMED AS E. COLI. THE SURGEON REVISED ALL MEDACTA COMPONENTS AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS ARE AVAILABLE; EXPLANTS ARE NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366257 VERSAFITCUP ACETABULAR SHELL Ø 46 CEMENTLESS ACETABULAR SHELL MEH MEDACTA INTERNATIONAL SA 153605

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention