25 G X 1 1/2 IN. BD ECLIPSE¿ NEEDLE
Report
- Report Number
- 8041187-2017-00044
- Event Type
- Malfunction
- Date Received
- May 22, 2017
- Date of Event
- April 12, 2017
- Report Date
- July 29, 2017
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FMI
- PMA / PMN Number
- K161170
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
RESULTS: 126 UNUSED REPRESENTATIVE SAMPLES IN SEALED PACKAGES WERE RETURNED FOR EVALUATION. THE SAMPLES WERE EVALUATED AND THE SAFETY SHIELDS DID NOT FALL OFF. CONCLUSION: AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE.
RESULTS: IT IS UNKNOWN IF A SAMPLE WILL BE RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6079229. A MANUFACTURING REVIEW REVEALED NO ABNORMALITIES WITH PREVENTATIVE MAINTENANCE, CALIBRATION, OR EQUIPMENT THAT COULD HAVE INFLUENCED THE CUSTOMER'S REPORTED ISSUE. IN THE EVENT THAT NEW, CHANGED, OR CORRECTED INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE FILED. CONCLUSION: WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE. REMEDIAL ACTION REQUIRED: CAPAS (B)(4) WERE OPENED TO IDENTIFY AND ADDRESS THE POTENTIAL CAUSES OF SAFETY SHIELD DISENGAGEMENT. ADDITIONALLY, FIELD ACTION NOTIFICATION MSS-16-837-FA WAS INITIATED AND A PRODUCT ADVISORY LETTER WAS SENT ON 12/29/2016.
IT WAS REPORTED THAT THE PINK SAFETY SHIELD OF A 25 G X 1 1/2 IN. BD ECLIPSE¿ NEEDLE FELL OFF DURING USE. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 362169 | 25 G X 1 1/2 IN. BD ECLIPSE¿ NEEDLE | SAFETY ENGINEERED HYPODERMIC NEEDLE | FMI | BECTON DICKINSON MEDICAL (SINGAPORE) | 6079229 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |