FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 6581448 · Received May 22, 2017

Report

Report Number
1220908-2017-01149
Event Type
Malfunction
Date Received
May 22, 2017
Date of Event
April 26, 2017
Report Date
May 2, 2017
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
DRT
UDI-DI
00847946005979
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE REPORTED MALFUNCTION WAS OBSERVED DURING REVIEW OF THE DEVICE'S ACTIVITY LOG. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING WITHOUT DUPLICATING THE REPORTED MALFUNCTION. THE DEVICE'S MONITOR BOARD WAS REPLACED AS A PRECAUTION. THE DEVICE SUCCESSFULLY PASSED THE FINAL TEST PROCEDURE, WAS RECERTIFIED, AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED AN"ECG DISABLED" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1220908-2017-01148 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
364359 X SERIES DEFIBRILLATOR/PACEMAKER DRT ZOLL MEDICAL CORPORATION X SERIES NA 00847946005979

Patients

Seq Age Sex Outcome Treatment
1