X SERIES
Report
- Report Number
- 1220908-2017-01149
- Event Type
- Malfunction
- Date Received
- May 22, 2017
- Date of Event
- April 26, 2017
- Report Date
- May 2, 2017
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- DRT
- UDI-DI
- 00847946005979
- PMA / PMN Number
- K112432
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE REPORTED MALFUNCTION WAS OBSERVED DURING REVIEW OF THE DEVICE'S ACTIVITY LOG. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING WITHOUT DUPLICATING THE REPORTED MALFUNCTION. THE DEVICE'S MONITOR BOARD WAS REPLACED AS A PRECAUTION. THE DEVICE SUCCESSFULLY PASSED THE FINAL TEST PROCEDURE, WAS RECERTIFIED, AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED AN"ECG DISABLED" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION. PLEASE REFERENCE MEDWATCH REPORT 1220908-2017-01148 FOR A SIMILAR EVENT REPORTED FROM THE SAME CUSTOMER.