INFUSOMAT®
Report
- Report Number
- 9610825-2017-00087
- Event Type
- Malfunction
- Date Received
- May 19, 2017
- Report Date
- April 21, 2017
- Manufacturer
- B. BRAUN MELSUNGEN AG
- Product Code
- FRN
- UDI-DI
- 04046963716752
- PMA / PMN Number
- K083689
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
EXEMPTION NUMBER E2016018 B. BRAUN INC. (BBMI) (IMPORTER) IS SUBMITTING THE REPORT ON BEHALF OF B. BRAUN MELSUNGEN AG (MANUFACTURER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER(B)(4). FAILURE ANALYSIS AND INVESTIGATION RESULTS DID NOT CONFIRM THE REPORTED ISSUE. UPON RECEIPT THE ACTUAL PUMP INVOLVED WAS VISUALLY AND FUNCTIONALLY INSPECTED. THREE (3) VOLUMETRICS OF 250ML/HR AND 25ML TESTED IN SPEC WITH NO FREE FLOW DOOR CLOSED OR OPENED. DURING THE EVALUATION OF THE PUMP, IT WAS DISCOVERED THAT THE PUMP HAD NO CONNECTION TO HIBASED, AND THE MOTHERBOARD WAS REPLACED. THIS ISSUE WOULD NOT HAVE RESULTED IN FREE FLOW, IT WOULD ONLY AFFECT THE PUMP'S ABILITY TO CONNECT TO THE FACILITY'S DRUG LIBRARY WIRELESSLY. REVIEW OF THE DISCREPANCY MANAGEMENT SYSTEM DATABASE PERFORMED FOR THE REPORTED LOT NUMBER DID NOT REVEAL ANY ABNORMALITIES OR NONCONFORMANCES OF THIS NATURE. NO ADVERSE QUALITY TRENDS OF THIS NATURE WERE IDENTIFIED DURING THE COMPLAINT REVIEW PROCESS FOR THE REPORTED CATALOG NUMBER. THE PUMP OPERATED AS INTENDED AND THE REPORTED FAILURE COULD NOT BE REPRODUCED. BASED ON THE RESULTS OF THE INVESTIGATION, NO CONCLUSIONS CAN BE MADE REGARDING THE CAUSE OF THE REPORTED EVENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
AS REPORTED BY USER FACILITY: FREE FLOW OF MAGNESIUM. PUMP DID NOT ALARM, THE ISSUE WAS SEEN BY A NURSE. NO PATIENT INJURY. THE INFUSION WAS STOPPED, AND RESTARTED ON ANOTHER PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359858 | INFUSOMAT® | PUMP, INFUSION | FRN | B. BRAUN MELSUNGEN AG | 04046963716752 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |