FDA Adverse Event Malfunction Summary report: N

INFUSOMAT®

MDR report key: 6579306 · Received May 19, 2017

Report

Report Number
9610825-2017-00087
Event Type
Malfunction
Date Received
May 19, 2017
Report Date
April 21, 2017
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
FRN
UDI-DI
04046963716752
PMA / PMN Number
K083689
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016018 B. BRAUN INC. (BBMI) (IMPORTER) IS SUBMITTING THE REPORT ON BEHALF OF B. BRAUN MELSUNGEN AG (MANUFACTURER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER(B)(4). FAILURE ANALYSIS AND INVESTIGATION RESULTS DID NOT CONFIRM THE REPORTED ISSUE. UPON RECEIPT THE ACTUAL PUMP INVOLVED WAS VISUALLY AND FUNCTIONALLY INSPECTED. THREE (3) VOLUMETRICS OF 250ML/HR AND 25ML TESTED IN SPEC WITH NO FREE FLOW DOOR CLOSED OR OPENED. DURING THE EVALUATION OF THE PUMP, IT WAS DISCOVERED THAT THE PUMP HAD NO CONNECTION TO HIBASED, AND THE MOTHERBOARD WAS REPLACED. THIS ISSUE WOULD NOT HAVE RESULTED IN FREE FLOW, IT WOULD ONLY AFFECT THE PUMP'S ABILITY TO CONNECT TO THE FACILITY'S DRUG LIBRARY WIRELESSLY. REVIEW OF THE DISCREPANCY MANAGEMENT SYSTEM DATABASE PERFORMED FOR THE REPORTED LOT NUMBER DID NOT REVEAL ANY ABNORMALITIES OR NONCONFORMANCES OF THIS NATURE. NO ADVERSE QUALITY TRENDS OF THIS NATURE WERE IDENTIFIED DURING THE COMPLAINT REVIEW PROCESS FOR THE REPORTED CATALOG NUMBER. THE PUMP OPERATED AS INTENDED AND THE REPORTED FAILURE COULD NOT BE REPRODUCED. BASED ON THE RESULTS OF THE INVESTIGATION, NO CONCLUSIONS CAN BE MADE REGARDING THE CAUSE OF THE REPORTED EVENT. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

AS REPORTED BY USER FACILITY: FREE FLOW OF MAGNESIUM. PUMP DID NOT ALARM, THE ISSUE WAS SEEN BY A NURSE. NO PATIENT INJURY. THE INFUSION WAS STOPPED, AND RESTARTED ON ANOTHER PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359858 INFUSOMAT® PUMP, INFUSION FRN B. BRAUN MELSUNGEN AG 04046963716752

Patients

Seq Age Sex Outcome Treatment
1