FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 657762 · Received December 29, 2005

Report

Report Number
2243072-2005-00519
Event Type
Other
Date Received
December 29, 2005
Date of Event
December 1, 2005
Report Date
December 29, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT ERRATIC READINGS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG READING (146; 188) AS REFERENCE POINT VS. BD READINGS (101, 273, 91,117, 94, 106, 432, 118) YIELDED SOME DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064223

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other