FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 6571042 · Received May 16, 2017

Report

Report Number
1314492-2017-00918
Event Type
Malfunction
Date Received
May 16, 2017
Date of Event
March 21, 2017
Report Date
May 16, 2017
Manufacturer
BAXTER HEALTHCARE - MEDINA
Product Code
FRN
PMA / PMN Number
K133801
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

BAXTER RECEIVED AND EVALUATED THE DEVICE. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A DEVICE HISTORY REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE AND DID NOT REVEAL ANY EVIDENCE OF NONCONFORMING PRODUCT. THE REPORTED CONDITION WAS VERIFIED. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYSTEM ERROR 324 ALARMS WHICH WERE REPRODUCED UPON POWERING UP THE DEVICE. SYSTEM ERROR 324 ALARMS WERE VERIFIED THROUGH A REVIEW OF THE EVENT HISTORY LOG AND FOUND TO BE CAUSED BY A FAILED UPPER AUXILIARY ASSEMBLY. THE UPPER AUXILIARY ASSEMBLY WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP EXPERIENCED A SYSTEM ERROR 324 (COLOR SENSOR ERROR: FRQ) ALARM. THERE WAS NO KNOWN PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
352376 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE - MEDINA N/A N/A

Patients

Seq Age Sex Outcome Treatment
1