FDA Adverse Event Malfunction Summary report: N

GUNTHER TULIP JUGULAR & FEMORAL VENA CAVA FILTER SET

MDR report key: 6568000 · Received May 15, 2017

Report

Report Number
1820334-2017-01212
Event Type
Malfunction
Date Received
May 15, 2017
Report Date
August 26, 2017
Manufacturer
COOK INC
Product Code
DTK
UDI-DI
10827002213606
PMA / PMN Number
K072240
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION. WILLIAM COOK EUROPE INITIALLY REPORTED EVENT UNDER MFR REPORT # 3002808486-2017-00699. NEW INFORMATION WAS RECEIVED IDENTIFYING THAT THE PRODUCT WAS A COOK INC.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: INVESTIGATION - FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. FILTER TILT IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE AND MAY OCCUR DURING PLACEMENT OR DURING IMPLANTING PERIOD. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 1

PLAINTIFF ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2012 DUE TO DEEP VEIN THROMBOSIS AND PULMONARY EMBOLISM. PLAINTIFF ALLEGES ATTEMPTED RETRIEVAL ON (B)(6) 2012. PLAINTIFF IS ALLEGING TILT, DEVICE UNABLE TO BE RETRIEVED. ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348866 GUNTHER TULIP JUGULAR & FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK COOK INC N/A 10827002213606

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening