FDA Adverse Event
Injury
Summary report: N
UNIVERSAL VENTILATOR CIRCUIT WITH NEBULIZER
MDR report key: 65679
·
Received January 31, 1997
Report
- Report Number
- 65679
- Event Type
- Injury
- Date Received
- January 31, 1997
- Date of Event
- January 16, 1997
- Report Date
- January 17, 1997
- Manufacturer
- PROFESSIONAL MEDICAL CIRCUIT WITH NEBULIZER
- Product Code
- BZO
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
BETWEEN THE HRS OF 4 AM AND 8 AM ON 1/16/97, FIVE LONG-TERM VENTILATOR PTS EXPERIENCED PROBLEMS WITH THEIR AIR-FLOW AND SUBSEQUENTLY DEVELOPED RESPIRATORY DISTRESS. AFTER STABILIZATION OF THESE PTS, AN INTERNAL INVESTIGATION WAS STARTED. THE FOLLOWING INFO WAS DISCOVERED: NEW VENTILATOR CIRCUITS WITH NEBULIZER HAD BEEN INSTALLED ON THE VENTS ON 1/12/97. THE LITERATURE THAT ACCOMPANIES THESE CIRCUITS STATES THAT THIS SYSTEM CAN BE USED WITH THE BRAND AND MODEL VENTILATOR THE FACILITY WAS USING. THE VENT CIRCUITS CAME EQUIPPED WITH AN "EXTERNAL EXHALATION VALVE" ATTACHED TO THE TUBING. THE VENTS CURRENTLY IN USE AT THIS MED CTR HAVE AN INTERNAL EXHALATION VALVE ALREADY IN PLACE. THIS IS PT 2 OF 5.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNIVERSAL VENTILATOR CIRCUIT WITH NEBULIZER | VENTILATOR CIRCUIT | BZO | PROFESSIONAL MEDICAL CIRCUIT WITH NEBULIZER | 201-5847 | 620091 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |