FDA Adverse Event Malfunction Summary report: N

GUNTHER TULIP FEMORAL VENA CAVA FILTER SET

MDR report key: 6562724 · Received May 12, 2017

Report

Report Number
1820334-2017-00864
Event Type
Malfunction
Date Received
May 12, 2017
Report Date
April 11, 2019
Manufacturer
COOK INC
Product Code
DTK
PMA / PMN Number
K043509
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: WE HAVE INVESTIGATED BASED ON THE INFORMATION RECEIVED TO DATE, AND ARE CLOSING THE REPORT UNTIL FURTHER INFORMATION IS RECEIVED FOR INVESTIGATION. NO CONCLUSION CAN BE DRAWN BASED ON THE INCOMPLETE INFORMATION PROVIDED ON ALLEGED INJURY. IF ADDITIONAL INFORMATION IS RECEIVED, THE REPORT WILL BE RE-OPENED FOR FURTHER INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION.

Additional Manufacturer Narrative · 1

CORRECTED DATA: ADVERSE EVENT AND PRODUCT PROBLEM TO PRODUCT PROBLEM. OTHER TO BLANK. ADDITIONAL INFORMATION: 510 (K) K043509. INVESTIGATION- INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'TULIP, TILT, VC PERFORATION'. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED. VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. FILTER TILT IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE AND MAY OCCUR DURING PLACEMENT OR DURING IMPLANTING PERIOD. NO RELEVANT NOTES FOUND IN DEVICE WORK ORDER. THERE IS ONE OTHER COMPLAINT ON LOT 2082981 RELATED TO VENA CAVA PERFORATION, HOWEVER THERE IS NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. PRODUCT IS MANUFACTURED AND INSPECTED ACCORDING TO MANUFACTURING INSTRUCTIONS AND QUALITY CONTROL. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Additional Manufacturer Narrative · 1

PER QUALITY ENGINEERING REVIEW, THE ADDITIONAL INFORMATION PROVIDED FOR THIS COMPLAINT DOES NOT CHANGE THE PREVIOUS INVESTIGATION CONCLUSION. THEREFORE, NO NEW INVESTIGATION ACTIVITIES WILL BE CONDUCTED AT THIS TIME. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED.

Description of Event or Problem · 1

IT IS ALLEGED THAT [PT] RECEIVED A COOK GUNTHER TULIP ON (B)(6) 2008. IT IS ALLEGED THAT THE PATIENT WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION RECEIVED ON 12MAY2017 ON: [PT] ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2008 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO DEEP VEIN THROMBOSIS WITH PULMONARY EMBOLISM. [PT] IS ALLEGING TILT, VENA CAVA PERFORATION.

Description of Event or Problem · 1

PER REVIEW OF ABDOMINAL/PELVIS CT, DATED (B)(4) 2017, "NO INFORMATION CONCERNING THE TYPE OF FILTER NOR DATE DEPLOYED. POSITIVE FOR CAVAL PERFORATION. SUPERIOR EXTENT IS AT THE INFERIOR BODY OF L2. INFERIOR EXTENT IS AT THE BODY OF L3. ITS SUPERIOR EXTENT JUST ABOVE THE RIGHT RENAL VEIN. CAVAL PERFORATION APPROXIMATELY 5 MM POSTERIORLY. THE CAVAL FILTER IS TILTED APPROXIMATELY 20 DEGREES ANTERIOR-TO-POSTERIOR IN THE SUPERIOR-TO-INFERIOR DIRECTION."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344594 GUNTHER TULIP FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK COOK INC N/A

Patients

Seq Age Sex Outcome Treatment
1 Other