FDA Adverse Event Death Summary report: N

BACT/ALERT 3D MICROBIAL DETECTION SYSTEM

MDR report key: 655758 · Received December 16, 2005

Report

Report Number
3002769706-2005-00035
Event Type
Death
Date Received
December 16, 2005
Date of Event
November 20, 2005
Report Date
December 13, 2005
Manufacturer
BIOMERIEUX, INC.
Product Code
JTA
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

THE CUSTOMER CALLED CO IN 2005 FOR ASSISTANCE IN CLEARING A STRING SPACE FORMAT ERROR ON THEIR BACT/VIEW SOFTWARE. WHEN A STRING SPACE FORMAT ERROR OCCURS, THE BACT/VIEW COMPUTER WILL STOP COMMUNICATING WITH A LIS SYSTEM UNTIL THE ERROR IS CLEARED. THE CUSTOMER INFORMED CO DURING THIS CALL OF A POSITIVE BACT/ALERT BOTTLE RESULT THAT WAS NOT TRANSMITTED TO THEIR LIS DURING THE TIME BEFORE THE ERROR MESSAGE WAS CLEARED. THE LAB WAS NOT STAFFED ON SUNDAY, AND THEY HAD TURNED OFF THEIR SYSTEM'S AUDIBLE ALARMS. THEY DID NOT REALIZE THAT A BOTTLE HAD GONE POSITIVE, AND THAT THE BACT/VIEW COMPUTER WAS NO LONGER COMMUNICATING WITH THEIR LIS SYSTEM UNTIL LATE MORNING OF THE SAME DAY. THE CUSTOMER CALLED CO BACK THE NEXT DAY IN THE EVENING TO INFORM CO THAT THE PT WHO RECEIVED THE CONTAMINATED PLATELETS HAD DIED. THE BLOOD BANK DETERMINED THAT THE PLATELETS WERE CONTAMINATED WITH E-COLI.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BACT/ALERT 3D MICROBIAL DETECTION SYSTEM MICROBIAL GROWTH MONITOR JTA BIOMERIEUX, INC. 210147 NA

Patients

Seq Age Sex Outcome Treatment
1 64 YR Death