SPACELABS ULTRAVIEW SL CENTRAL MONITOR

MDR report key: 6557390 · Received May 10, 2017

Report

Report Number
3010157426-2017-00041
Event Type
Injury
Date Received
May 10, 2017
Date of Event
April 3, 2017
Report Date
May 10, 2017
Manufacturer
SPACELABS HEALTHCARE INC.
Product Code
MHX
PMA / PMN Number
K102422
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVENT, STAFF CONFIRM OTHER BEDSIDE ALARMS WERE OCCURRING AT THE CENTRAL MONITOR AND THE CUSTOMER BIOMED RESTORED ALARMS AT THE CENTRAL FOR THE BEDSIDE DEVICE IN QUESTION BY PERFORMING A SCREEN FORMAT. SPACELABS WAS NOTIFIED A WEEK AFTER THE EVENT OCCURRED, AND AN INVESTIGATION WAS IMMEDIATELY LAUNCHED. HOWEVER, DUE TO THE LENGTH OF TIME THAT HAD PASSED, THE PATIENT'S HISTORICAL DATA AND TREND INFORMATION WAS NO LONGER AVAILABLE. A SPACELABS FIELD SERVICE ENGINEER ARRIVED TO INVESTIGATE ONSITE. CLINICAL STAFF STATED ALARMS WERE HEARD DURING THE EVENT FROM THE BEDSIDE MONITOR, HOWEVER A DECISION WAS MADE BY THE CUSTOMER TO REMOVE THE TWENTY ONE YEAR OLD BEDSIDE MONITOR FROM SERVICE FOLLOWING THE EVENT. BIOMEDICAL STAFF DECLINED TO PROVIDE APPROVAL TO TEST THE INVOLVED BEDSIDE AND CENTRAL MONITOR, THEREFORE SPACELABS IS UNABLE TO INVESTIGATE ANY FURTHER. THE BEDSIDE MONITOR IS CONSIDERED THE PRIMARY ALARM NOTIFICATION DEVICE. CONFIRMATION WAS RECEIVED THAT THE BEDSIDE MONITOR DID GENERATE ALARMS DURING THE EVENT. THIS INVESTIGATION IS CONSIDERED COMPLETE AND THE MATTER CLOSED.

Description of Event or Problem · 1

SPACELABS RECEIVED A REPORT THAT ON (B)(6) 2017, AT 05:38 AM, AN ALARM DID NOT OCCUR AT THE CENTRAL MONITOR DURING A CARDIAC EVENT INVOLVING A PATIENT ON A BEDSIDE MONITOR. ALARMS GENERATED FROM THE PATIENT¿S BEDSIDE MONITOR. THE PATIENT REMAINED HOSPITALIZED AND DIED TWO DAYS LATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339247 SPACELABS ULTRAVIEW SL CENTRAL MONITOR UVSL CENTRAL MONITOR MHX SPACELABS HEALTHCARE INC. 91387-38

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death 90385 ((B)(4))