FDA Adverse Event Malfunction Summary report: N

CALIFORNIA MEDICAL LABORATORIES INC MALLEABLE VENT CATHETER

MDR report key: 6556902 · Received May 10, 2017

Report

Report Number
1718850-2017-00007
Event Type
Malfunction
Date Received
May 10, 2017
Report Date
July 10, 2017
Manufacturer
SORIN GROUP USA, INC
Product Code
DWF
UDI-DI
00803622106439
PMA / PMN Number
K981601
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A PHOTOGRAPH OF AN UNUSED MALLEABLE VENT CATHETER WAS PROVIDED BY THE CUSTOMER, SHOWING THE AREA OF DISCONNECTION TO BE AT THE CATHETER TO VENT CONNECTION JUNCTION. RESEARCH INTO THE REPORTED LOT NUMBER FOUND THAT THERE WAS NO REMAINING INVENTORY AVAILABLE FOR FURTHER INVESTIGATION. THERE HAVE BEEN OVER 110 BUILDS OF THIS CATALOG NUMBER WITHIN THE PAST YEAR, REPRESENTING ALMOST 20,000 UNITS. THIS IS THE FIRST REPORT AGAINST THIS CATALOG NUMBER FOR ANY REASON, INCLUDING DISCONNECTION. WITHOUT THE ABILITY TO CONDUCT AN IN-DEPTH EVALUATION OF THE PRODUCT, A ROOT CAUSE COULD NOT BE DETERMINED. TO ADDRESS ANY POSSIBLE PRODUCTION ASSEMBLY ISSUES, A MANUFACTURING AWARENESS NOTIFICATION WAS ISSUED TO THE APPROPRIATE WORKFORCE, ALERTING AND SENSITIZING THE WORKFORCE TO THE REPORTED MALFUNCTION. PHOTO RECEIVED; ACTUAL DEVICE DISCARDED.

Additional Manufacturer Narrative · 1

THROUGH FOLLOW-UP COMMUNICATION WITH THE CUSTOMER, SORIN GROUP LEARNED THAT THE DEVICE HAS BEEN DISCARDED BY THE CUSTOMER. IT WAS ALSO REPORTED THAT THERE WAS NO DELAY IN THE SURGERY. THE INVESTIGATION IS ON-GOING. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. DEVICE DISCARDED BY THE CUSTOMER.

Description of Event or Problem · 1

SORIN GROUP RECEIVED A REPORT THAT THE TIP OF THE MALLEABLE VENT CATHETER DISCONNECTED DURING A PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340232 CALIFORNIA MEDICAL LABORATORIES INC MALLEABLE VENT CATHETER CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS DWF SORIN GROUP USA, INC 1633600160 00803622106439

Patients

Seq Age Sex Outcome Treatment
1