FDA Adverse Event Injury Summary report: N

FMC- DIALYSIS MACHINE

MDR report key: 6555480 · Received March 17, 2017

Report

Report Number
6555480
Event Type
Injury
Date Received
March 17, 2017
Date of Event
December 30, 2016
Report Date
January 12, 2017
Product Code
KDI
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

DURING THE 3RD SHIFT ON B-SIDE OF THE OP DIALYSIS UNIT, 3 PATIENTS WERE FINISHED AND 5 REMAINED. THE 5 WERE HALFWAY DONE WITH THEIR TREATMENT WHEN THEIR MACHINES BEGAN ALARMING SPORADICALLY AND PATIENTS BEGAN TO COMPLAIN OF TINGLING IN HANDS AND FACE. SOME EXPERIENCED MODERATE HYPOTENSION. TREATMENT STOPPED AND BLOOD WAS RETURNED. RENAL PANELS AND BLOOD CULTURES WERE OBTAINED. PATIENTS WERE GIVEN NOVASOURCE. AFTER LABS WERE RESULTED, THE PATIENTS WERE MOVED TO A SIDE FOR ONE HOUR HEMODIALYSIS WITH A 3K BATH. SYMPTOMS RESOLVED AND PATIENTS WERE DISCHARGED. THIS PATIENT'S SYMPTOMS WERE TINGLING TO HANDS AND FACE. SHE WAS IN CHAIR 17. 2K BATH. UNITS A AND B ARE ON DIFFERENT LOOPS. DAILY BICARB AND ACID ADDED AT 4:30AM FOLLOWING ROUTINE PROCESS AND USUAL STAFF. RO WATER TESTED NORMAL. STORAGE TANKS FOR BICARB CHECKED AFTER INCIDENT. NO SLUDGE SEEN. CONFERENCE CALL TO FRESENIUS ON (B)(6) 2017 TO DISCUSS INCIDENT AND ASK FOR ASSISTANCE IN DETERMINING POSSIBLE CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197937 FMC- DIALYSIS MACHINE FMC- DIALYSIS MACHINE KDI

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other