FDA Adverse Event Malfunction Summary report: N

ARCHITECT SHBG

MDR report key: 6552913 · Received May 8, 2017

Report

Report Number
3002809144-2017-00034
Event Type
Malfunction
Date Received
May 8, 2017
Date of Event
April 10, 2017
Report Date
May 8, 2017
Manufacturer
ABBOTT GERMANY
Product Code
CDZ
PMA / PMN Number
K060818
Removal / Correction Number
3002809144-04/18/17-001-
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CORRECTION/REMOVAL REPORTING NUMBER: 3002809144-04/18/17-001-C WAS TOO LONG TO FIT IN SECTION. A PRODUCT CORRECTION LETTER WAS ISSUED TO ALL CUSTOMERS WHO RECEIVED ONE OR MORE OF THE ARCHITECT SHBG REAGENT LOTS LISTED BELOW. THE CUSTOMERS WERE INSTRUCTED TO IMMEDIATELY DISCONTINUE USE OF POTASSIUM EDTA SPECIMEN TUBE TYPES WITH THE ARCHITECT SHBG ASSAY. A KIT STUFFER WILL BE INCLUDED WITH ALL CURRENT INVENTORY ADVISING THE CUSTOMER TO DISCONTINUE THE USE OF POTASSIUM EDTA TUBE TYPES. THE REAGENT PACKAGE INSERT FOR ALL FUTURE LOTS OF ARCHITECT SHBG HAS BEEN UPDATED TO REMOVE POTASSIUM EDTA TUBES AS AN ACCEPTABLE TUBE TYPE FOR SPECIMEN COLLECTION. ALL IN-DATE ARCHITECT SHBG LOT NUMBERS ARE IMPACTED AS FOLLOWS: LIST NUMBER, LOT NUMBER, MANUFACTURE DATE, EXPIRATION DATE 8K26-20, 01916E000, 13-JUN-2016, 25-APR-2017 8K26-20, 00316F000, 04-JUL-2016, 27-MAY-2017 8K26-20, 01816G000, 31-AUG-2016, 12-JUL-2017 8K26-20, 00916I000, 05-OCT-2016, 05-SEP-2017 8K26-20, 07316I000, 07-DEC-2016, 21-OCT-2017 8K26-20, 00916L000, 20-JAN-2017, 16-DEC-2017 8K26-25, 01816E000, 13-JUN-2016, 25-APR-2017 8K26-25, 00216F000, 04-JUL-2016, 27-MAY-2017 8K26-25, 01716G000, 31-AUG-2016, 12-JUL-2017 8K26-25, 00816I000, 05-OCT-2016, 05-SEP-2017 8K26-25, 07216I000, 07-DEC-2016, 21-OCT-2017

Description of Event or Problem · 1

ABBOTT LABORATORIES HAS IDENTIFIED A POTENTIAL FOR FALSELY DECREASED ARCHITECT SHBG RESULTS TO BE GENERATED WHEN USING THE ARCHITECT SHBG ASSAY WITH THE POTASSIUM EDTA SPECIMEN TUBE TYPE. THIS MAY ALSO RESULT IN AN INCREASE IN FREE ANDROGEN INDEX / FREE TESTOSTERONE INDEX CALCULATIONS OF UP TO 30% WHEN USING POTASSIUM EDTA SPECIMEN TUBES.NO PATIENT INJURY HAS BEEN REPORTED DUE TO THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332932 ARCHITECT SHBG TESTOSTERONE TEST SYSTEM CDZ ABBOTT GERMANY

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT ISYSTEM ANALYZER