ARCHITECT SHBG
Report
- Report Number
- 3002809144-2017-00034
- Event Type
- Malfunction
- Date Received
- May 8, 2017
- Date of Event
- April 10, 2017
- Report Date
- May 8, 2017
- Manufacturer
- ABBOTT GERMANY
- Product Code
- CDZ
- PMA / PMN Number
- K060818
- Removal / Correction Number
- 3002809144-04/18/17-001-
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
CORRECTION/REMOVAL REPORTING NUMBER: 3002809144-04/18/17-001-C WAS TOO LONG TO FIT IN SECTION. A PRODUCT CORRECTION LETTER WAS ISSUED TO ALL CUSTOMERS WHO RECEIVED ONE OR MORE OF THE ARCHITECT SHBG REAGENT LOTS LISTED BELOW. THE CUSTOMERS WERE INSTRUCTED TO IMMEDIATELY DISCONTINUE USE OF POTASSIUM EDTA SPECIMEN TUBE TYPES WITH THE ARCHITECT SHBG ASSAY. A KIT STUFFER WILL BE INCLUDED WITH ALL CURRENT INVENTORY ADVISING THE CUSTOMER TO DISCONTINUE THE USE OF POTASSIUM EDTA TUBE TYPES. THE REAGENT PACKAGE INSERT FOR ALL FUTURE LOTS OF ARCHITECT SHBG HAS BEEN UPDATED TO REMOVE POTASSIUM EDTA TUBES AS AN ACCEPTABLE TUBE TYPE FOR SPECIMEN COLLECTION. ALL IN-DATE ARCHITECT SHBG LOT NUMBERS ARE IMPACTED AS FOLLOWS: LIST NUMBER, LOT NUMBER, MANUFACTURE DATE, EXPIRATION DATE 8K26-20, 01916E000, 13-JUN-2016, 25-APR-2017 8K26-20, 00316F000, 04-JUL-2016, 27-MAY-2017 8K26-20, 01816G000, 31-AUG-2016, 12-JUL-2017 8K26-20, 00916I000, 05-OCT-2016, 05-SEP-2017 8K26-20, 07316I000, 07-DEC-2016, 21-OCT-2017 8K26-20, 00916L000, 20-JAN-2017, 16-DEC-2017 8K26-25, 01816E000, 13-JUN-2016, 25-APR-2017 8K26-25, 00216F000, 04-JUL-2016, 27-MAY-2017 8K26-25, 01716G000, 31-AUG-2016, 12-JUL-2017 8K26-25, 00816I000, 05-OCT-2016, 05-SEP-2017 8K26-25, 07216I000, 07-DEC-2016, 21-OCT-2017
ABBOTT LABORATORIES HAS IDENTIFIED A POTENTIAL FOR FALSELY DECREASED ARCHITECT SHBG RESULTS TO BE GENERATED WHEN USING THE ARCHITECT SHBG ASSAY WITH THE POTASSIUM EDTA SPECIMEN TUBE TYPE. THIS MAY ALSO RESULT IN AN INCREASE IN FREE ANDROGEN INDEX / FREE TESTOSTERONE INDEX CALCULATIONS OF UP TO 30% WHEN USING POTASSIUM EDTA SPECIMEN TUBES.NO PATIENT INJURY HAS BEEN REPORTED DUE TO THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332932 | ARCHITECT SHBG | TESTOSTERONE TEST SYSTEM | CDZ | ABBOTT GERMANY |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT ISYSTEM ANALYZER |