MAKOPLASTY® PARTIAL KNEE END EFFECTOR
Report
- Report Number
- 3005985723-2017-00202
- Event Type
- Malfunction
- Date Received
- May 8, 2017
- Date of Event
- May 2, 2017
- Report Date
- July 9, 2017
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: THE EVENT REPORTED THAT A PARTIAL KNEE END EFFECTOR WAS MISSING ITS TRIGGER MAGNET. DEVICE EVALUATION AND RESULTS: UNABLE TO PERFORM AS THE PRODUCT WAS NOT AVAILABLE. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE THE FOLLOWING: (B)(4). QT 16-03-0076 DISPOSITIONED THE REJECTED PART AS RTV FOR WORKMANSHIP/DAMAGE. THE REJECTED PART WAS NOT THIS PART AND THE NONCONFORMANCE WAS UNRELATED TO THE ALLEGED FAILURE IN THIS INVESTIGATION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE RELATED TO P/N 111758, L/N 19330515 SHOWS TEN (10) COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. THE (B)(4). CONCLUSION: THE FAILURE MODE WAS CONFIRMED. MATERIAL ANALYSIS WAS PERFORMED ON OTHER DEVICES OF THE SAME PART NUMBER AND LOT AND RESULTS SHOWED THE PARTS HAD NO ADHESIVE REMAINS AT THE SIDEWALLS OF THE CAVITY AND VERY LITTLE REMNANTS AT THE BASE OF THE CAVITY FOR THOSE DEVICES THAT WERE UNDERWENT MAGNET DISASSOCIATION OR LOOSENING. CORRECTIVE ACTION / PREVENTIVE ACTION: (B)(4) HAS BEEN CLOSED FOR THE FAILURE MODE OF MAGNET DISASSOCIATION EXCEEDING THE ACCEPTABLE OCCURRENCE RATE. CAPA (B)(4) HAS BEEN INITIATED FOR THE FAILURE MODE.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
END EFFECTOR ASSEMBLY IS MISSING MAGNET IN ITS TRIGGER NECESSARY TO ACTIVATE BURR MOTOR.
END EFFECTOR ASSEMBLY IS MISSING MAGNET IN ITS TRIGGER NECESSARY TO ACTIVATE BURR MOTOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332675 | MAKOPLASTY® PARTIAL KNEE END EFFECTOR | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |