FDA Adverse Event Malfunction Summary report: N

HIP END EFFECTOR, VARIABLE ANGLE

MDR report key: 6552658 · Received May 8, 2017

Report

Report Number
3005985723-2017-00203
Event Type
Malfunction
Date Received
May 8, 2017
Date of Event
May 4, 2017
Report Date
June 2, 2017
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K141989
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT: WINGNUT ON BACK OF END EFFECTOR COMPLETELY CAME OFF AND WOULD NOT REATTACH TO END EFFECTOR. DELAY OF 5 MINUTES DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION CONFIRMS FAILED LOCKING MECHANISM WITH BALL BEARINGS FAILING. ALSO VISUAL INSPECTION CONFIRMS A SHEARED OFF 202866 HIP RELEASE KNOB. REFERENCE ATTACHED IMAGE. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED AS THE VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION WAS NOT COMPLETED AS THE VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON 03/31/2016 INCLUDING SERIAL NUMBER (B)(4). COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 206967, LOT 19020415 SHOWS 0 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: ALLEGED FAILURE CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA (B)(4) WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

WINGNUT ON BACK OF END EFFECTOR COMPLETELY CAME OFF AND WOULD NOT REATTACH TO END EFFECTOR. DELAY OF 5 MINUTES.

Description of Event or Problem · 1

WING NUT ON BACK OF END EFFECTOR COMPLETELY CAME OFF AND WOULD NOT REATTACH TO END EFFECTOR. DELAY OF 5 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332561 HIP END EFFECTOR, VARIABLE ANGLE STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP.

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization