HIP END EFFECTOR, VARIABLE ANGLE
Report
- Report Number
- 3005985723-2017-00203
- Event Type
- Malfunction
- Date Received
- May 8, 2017
- Date of Event
- May 4, 2017
- Report Date
- June 2, 2017
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K141989
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: WINGNUT ON BACK OF END EFFECTOR COMPLETELY CAME OFF AND WOULD NOT REATTACH TO END EFFECTOR. DELAY OF 5 MINUTES DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION: VISUAL INSPECTION CONFIRMS FAILED LOCKING MECHANISM WITH BALL BEARINGS FAILING. ALSO VISUAL INSPECTION CONFIRMS A SHEARED OFF 202866 HIP RELEASE KNOB. REFERENCE ATTACHED IMAGE. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED AS THE VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION WAS NOT COMPLETED AS THE VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON 03/31/2016 INCLUDING SERIAL NUMBER (B)(4). COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 206967, LOT 19020415 SHOWS 0 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: ALLEGED FAILURE CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: CAPA (B)(4) WAS INITIATED BASED ON OTHER COMPLAINTS REPORTING THIS ISSUE.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
WINGNUT ON BACK OF END EFFECTOR COMPLETELY CAME OFF AND WOULD NOT REATTACH TO END EFFECTOR. DELAY OF 5 MINUTES.
WING NUT ON BACK OF END EFFECTOR COMPLETELY CAME OFF AND WOULD NOT REATTACH TO END EFFECTOR. DELAY OF 5 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332561 | HIP END EFFECTOR, VARIABLE ANGLE | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |