BD NEOFLON¿ IV CANNULA SYSTEM 24 G X 1 IN.
Report
- Report Number
- 8041187-2017-00041
- Event Type
- Injury
- Date Received
- May 8, 2017
- Date of Event
- April 13, 2017
- Report Date
- May 30, 2017
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FOZ
- PMA / PMN Number
- K850276
- Removal / Correction Number
- 81917
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
RESULTS: TWO USED AND 48 UNUSED SAMPLES WERE RETURNED FOR EVALUATION. A VISUAL INSPECTION OF THE USED SAMPLES REVEALED CATHETER PEELBACK. TEN OF THE 48 UNUSED SAMPLES WERE TESTED FOR BEVEL ANGLE AND TIP OUTSIDE DIAMETER MEASUREMENTS. THE SAMPLES MET THE BEVEL ANGEL AND TIP OD SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6230410. A MANUFACTURING REVIEW OBSERVED NO ABNORMALITIES THAT COULD HAVE INFLUENCED THE REPORTED ISSUE. CONCLUSION: THE ROOT CAUSE FOR THIS INCIDENT WAS DETERMINED TO BE MATERIAL USED TO MANUFACTURE THE DEVICE. REMEDIAL ACTION REQUIRED: CAPA (B)(4) HAS BEEN INITIATED TO REVIEW THE IV CATHETER TUBING MATERIAL.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT THE CATHETER OF A BD NEOFLON¿ IV CANNULA SYSTEM 24 G X 1 IN. BENT AND TWISTED WHILE BEING INSERTED. THERE WAS NO HARM TO THE PATIENT BUT THE INCIDENT RESULTED IN A NEEDLE STICK TO THE MEDICAL PROFESSIONAL. IT WAS ALSO REPORTED THAT NO MEDICAL INTERVENTION WAS NECESSARY/PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332732 | BD NEOFLON¿ IV CANNULA SYSTEM 24 G X 1 IN. | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL (SINGAPORE) | 6230410 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |