FDA Adverse Event
Other
Summary report: N
LIFE-AIR 1000 HYPOTHERMIC THERAPY SYSTEM
MDR report key: 65496
·
Received December 26, 1996
Report
- Report Number
- 65496
- Event Type
- Other
- Date Received
- December 26, 1996
- Date of Event
- November 20, 1996
- Report Date
- December 17, 1996
- Manufacturer
- PROGRESSIVE DYNAMICS, INC.
- Product Code
- DWJ
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- NH, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
SECOND DEGREE THERMAL BURNS TO LEFT HEEL AND RIGHT GREAT TOE FOLLOWING EMERGENCY ABDOMINAL HYSTERECTOMY FOR POST-PARTUM BLEEDING. THE HOSE FROM A LIFE-AIR FORCED AIR WARMER WAS PLACED UNDER COTTON BLANKETS WITH THE OUTLET NEAR THE PT'S FEET DURING THE SURGERY. TEMP SETTING WAS 100 DEGREE F. NO WARMING COVER WAS USED. AT THE CONCLUSION OF SURGERY THE PT'S FEET WERE ERYTHEMATOUS AND PAINFUL AND A BLISTER WAS NOTED ON THE RIGHT GREAT TOE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFE-AIR 1000 HYPOTHERMIC THERAPY SYSTEM | HYPOTHERMIC THERAPY SYSTEM | DWJ | PROGRESSIVE DYNAMICS, INC. | PD 100-S | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Other |