FDA Adverse Event Other Summary report: N

LIFE-AIR 1000 HYPOTHERMIC THERAPY SYSTEM

MDR report key: 65496 · Received December 26, 1996

Report

Report Number
65496
Event Type
Other
Date Received
December 26, 1996
Date of Event
November 20, 1996
Report Date
December 17, 1996
Manufacturer
PROGRESSIVE DYNAMICS, INC.
Product Code
DWJ
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
NH, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

SECOND DEGREE THERMAL BURNS TO LEFT HEEL AND RIGHT GREAT TOE FOLLOWING EMERGENCY ABDOMINAL HYSTERECTOMY FOR POST-PARTUM BLEEDING. THE HOSE FROM A LIFE-AIR FORCED AIR WARMER WAS PLACED UNDER COTTON BLANKETS WITH THE OUTLET NEAR THE PT'S FEET DURING THE SURGERY. TEMP SETTING WAS 100 DEGREE F. NO WARMING COVER WAS USED. AT THE CONCLUSION OF SURGERY THE PT'S FEET WERE ERYTHEMATOUS AND PAINFUL AND A BLISTER WAS NOTED ON THE RIGHT GREAT TOE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFE-AIR 1000 HYPOTHERMIC THERAPY SYSTEM HYPOTHERMIC THERAPY SYSTEM DWJ PROGRESSIVE DYNAMICS, INC. PD 100-S NA

Patients

Seq Age Sex Outcome Treatment
1 34 YR Other