FDA Adverse Event
Malfunction
Summary report: N
XIO RTP SYSTEM
MDR report key: 654770
·
Received December 14, 2005
Report
- Report Number
- 1937649-2005-00002
- Event Type
- Malfunction
- Date Received
- December 14, 2005
- Date of Event
- November 22, 2005
- Report Date
- December 13, 2005
- Manufacturer
- COMPUTERIZED MEDICAL SYSTEMS, INC.
- Product Code
- IYE
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
A USER HAD DEVELOPED A TREATMENT PLAN USING THE XIO SOFTWARE AND NEEDED TO SWITCH FROM ONE ELECTRON MACHINE TO ANOTHER DUE TO RESTRICTIONS IN THE SSD RANGE. XIO DETECTED AN INVALID TREATMENT DISTANCE, AND THE "DOSE NOT CURRENT" MESSAGE WAS DISPLAYED AT THE BOTTOM OF THE XIO SCREEN, AS WAS APPROPRIATE. HOWEVER, WHEN THE USER ENTERED A VALID TREATMENT DISTANCE AND THE MESSAGE WAS REMOVED, XIO DID NOT RECALCULATE THE BEAM, AND THE DOSES FOR THE PREVIOUSLY CALCULATED MACHINE WERE REDISPLAYED. THE PROBLEM WAS DETECTED, AND NO PT'S WERE MISTREATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XIO RTP SYSTEM | RADIATION TREATMENT PLANNING SYSTEM | IYE | COMPUTERIZED MEDICAL SYSTEMS, INC. | AII | 4.1.0-4.3.0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |