FDA Adverse Event Malfunction Summary report: N

XIO RTP SYSTEM

MDR report key: 654760 · Received December 14, 2005

Report

Report Number
1937649-2005-00003
Event Type
Malfunction
Date Received
December 14, 2005
Date of Event
December 2, 2005
Report Date
December 14, 2005
Manufacturer
COMPUTERIZED MEDICAL SYSTEMS, INC.
Product Code
IYE
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
PHYSICIST

Narratives

Description of Event or Problem · 1

A USER HAD DEVELOPED A TREATMENT PLAN USING THE XIO SOFTWARE AND SWITCHED FROM A COBALT UNIT TO A LINEAR ACCELERATOR. XIO DETECTED AN INVALID TREATMENT DISTANCE, AND THE "DOSE NOT CURRENT" MESSAGE WAS DISPLAYED AT THE BOTTOM OF THE XIO SCREEN, AS WAS APROPRIATE. HOWEVER, WHEN THE USER ENTERED A VALID TREATMENT DISTANCE AND THE MESSAGE WAS REMOVED, XIO DID NOT RECALCULATE THE BEAM, AND THE DOSES FOR THE PREVIOUSLY CALCULATED MACHINE WERE REDISPLAYED. THE PROBLEM WAS DETECTED, AND NO PTS WERE MISTREATED. THIS SAME ISSUE WAS REPORTED UNDER MDR 1937649-2005-002. THIS MDR REPRESENTS A SECOND USER REPORT OF THE SAME SOFTWARE BUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XIO RTP SYSTEM RADIATION TREATMENT PLANNING SYSTEM IYE COMPUTERIZED MEDICAL SYSTEMS, INC. ALL 4.1.0 - 4.3.0

Patients

Seq Age Sex Outcome Treatment
1 NA