FDA Adverse Event
Malfunction
Summary report: N
PREMIER PRO
MDR report key: 6546073
·
Received May 4, 2017
Report
- Report Number
- 6546073
- Event Type
- Malfunction
- Date Received
- May 4, 2017
- Date of Event
- April 3, 2017
- Report Date
- April 11, 2017
- Manufacturer
- S2S GLOBAL
- Product Code
- KZD
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
1. PRESSURE BAGS OVER INFLATE (AFTER THEY ARE HANGING FOR A WHILE) AND CAUSE THE FLUID BAG TO RUPTURE AND SPRAY/LEAK. 2. PREMIER PRO 1 LITER PRESSURE BAGS (REORDER NUMBER (B)(4)) DO NOT ENCOMPASS THE ENTIRE LITER BAG OF SALINE. WHEN PUMPED UP TO 300 MMHG, THE SALINE BAGS ARE BURSTING. 3. THE PRESSURE GAUGE OFTEN DID NOT ENGAGE UNTIL THE BAG WAS FILLED/OVERFILLED AND THEN WOULD POP OUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324390 | PREMIER PRO | INFUSOR, PRESSURE, FOR I.V. BAGS | KZD | S2S GLOBAL | 8801 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |