FDA Adverse Event Malfunction Summary report: N

PREMIER PRO

MDR report key: 6546073 · Received May 4, 2017

Report

Report Number
6546073
Event Type
Malfunction
Date Received
May 4, 2017
Date of Event
April 3, 2017
Report Date
April 11, 2017
Manufacturer
S2S GLOBAL
Product Code
KZD
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

1. PRESSURE BAGS OVER INFLATE (AFTER THEY ARE HANGING FOR A WHILE) AND CAUSE THE FLUID BAG TO RUPTURE AND SPRAY/LEAK. 2. PREMIER PRO 1 LITER PRESSURE BAGS (REORDER NUMBER (B)(4)) DO NOT ENCOMPASS THE ENTIRE LITER BAG OF SALINE. WHEN PUMPED UP TO 300 MMHG, THE SALINE BAGS ARE BURSTING. 3. THE PRESSURE GAUGE OFTEN DID NOT ENGAGE UNTIL THE BAG WAS FILLED/OVERFILLED AND THEN WOULD POP OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324390 PREMIER PRO INFUSOR, PRESSURE, FOR I.V. BAGS KZD S2S GLOBAL 8801

Patients

Seq Age Sex Outcome Treatment
1 31 YR