FDA Adverse Event Other Summary report: N

AEROSET ANALYZER

MDR report key: 654331 · Received December 16, 2005

Report

Report Number
1628664-2005-00050
Event Type
Other
Date Received
December 16, 2005
Date of Event
November 23, 2005
Report Date
December 15, 2005
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER STATES THAT PT AEROSET ICT MODULE RESULTS GENERATE ACCEPTABLE RESULTS AT THE BEGINNING OF THE DAY BUT THEN BEGIN DRIFTING LOW OVER TIME. FOR EXAMPLE, ONE PT GENERATED RESULTS AS FOLLOWS: SODIUM = 119 MEQ/L, POTASSIUM = 3.8 MEQ/L, AND CHLORIDE = 107 MEQ/L. THE SAMPLE WAS THEN TESTED ON ANOTHER AEROSET ANALYZER IN THE LAB AND GENERATED THE FOLLOWING RESULTS: SODIUM = 144 MEQ/L, POTASSIUM = 3.8 MEQ/L, AND CHLORIDE = 107 MEQ/L. THE CUSTOMER STATES THAT NO SUSPECT RESULTS HAVE BEEN REPORTED FROM THE LAB AND THAT CONTROLS REMAIN WITHIN SPECIFICATIONS. THE CUSTOMER PERFORMED SEVERAL SUGGESTED TROUBLESHOOTING PROCEDURES AND PT RESULTS THEN BEGAN GENERATING HIGHER THAN EXPECTED RESULTS. AN INITIAL SERVICE CALL DID NOT RESOLVE THE ISSUE AND THE CUSTOMER IS REQUESTING FIELD SERVICE TO RETURN. THERE IS NO IMPACT TO PT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AEROSET ANALYZER AUTOMATED CHEMISTRY ANALYZER JJE ABBOTT MANUFACTURING, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN