TRANSPAC IV MONITORING KIT WITH SAFESET RESERVOIR
Report
- Report Number
- 2025816-2017-00112
- Event Type
- Malfunction
- Date Received
- May 3, 2017
- Date of Event
- February 15, 2016
- Report Date
- May 3, 2017
- Manufacturer
- ICU MEDICAL INC.
- Product Code
- DRS
- PMA / PMN Number
- K061573
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- NURSE
Narratives
LOT REVIEW: A REVIEW OF THE MFG. LOT DATABASE FOR SUSPECT LOT# 3330120 (MFG. DATE 09/2016) SHOWS (B)(4) UNITS WERE MFGD., TESTED, INSPECTED, AND RELEASED, CITING NO EXCEPTION DOCUMENTS. VISUAL INSPECTION: RECEIVED ONE USED 42646-06, TRANSPAC IV MONITORING KIT. THE RESERVOIR AND TWO SSII PORTS RETURNED AND 12 INCH TUBING. THE MALE LUER CONNECTOR ON THE 3 INCH TUBING WAS SEPARATED AT THE BOND. THERE IS A SOLVENT RING ON THE TUBING. THE PRESSURE TUBING OF THE 3" DISTAL SECTION AND THE DISTAL MALE LUER WAS VISUALLY OBSERVED UNDER MAGNIFICATION. THE PRESSURE TUBING AND THE DISTAL MALE LUER WERE OBSERVED TO HAVE A FULL AND COMPLETE SOLVENT RING. FINAL ANALYSIS SUMMARY: THE REPORTED COMPLIANT OF BROKEN WAS VISUALLY CONFIRMED. THE ROOT CAUSE OF THE FAILURE IS BELIEVED TO HAVE BEEN CAUSED FROM UNINTENDED FORCE. THE TUBING AND THE LUER BOTH EXHIBITED PROMINENT SOLVENT RINGS. THESE RINGS INDICATE THAT THE BOND WAS ADEQUATE. ICU MEDICAL WILL CONTINUE TO MONITOR AND TREND THE REPORTED COMPLAINT.
(B)(6) COMPLAINT REC'D REGARDING ONE (1) 42646-06, TRANSPAC IV MONITORING KIT WITH SAFESET RESERVOIR AND 2 BLOOD SAMPLING PORTS, 84" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT); LOT# UNKNOWN. THE REPORT STATES, THE PATIENT NOTIFIED THE NURSE THAT THERE WAS A PROBLEM. UPON INSPECTION, THE NURSE REALIZED THAT THE PLASTIC HAD BROKEN AND WAS NOT AT AN CONNECTION POINT. NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322029 | TRANSPAC IV MONITORING KIT WITH SAFESET RESERVOIR | TRANSPAC IV MONITORING KIT WITH SAFESET RESERVOIR | DRS | ICU MEDICAL INC. | 42646-06 | 3330120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |