FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 6537087 · Received May 2, 2017

Report

Report Number
2531779-2017-09125
Event Type
Injury
Date Received
May 2, 2017
Date of Event
March 22, 2017
Report Date
April 5, 2017
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/17/2017 WITH THE FOLLOWING FINDINGS: THE EVENT DATE OF (B)(6) 2017 WAS NOT AVAILABLE IN THE BLACK BOX. THE BLACK BOX SHOWED A MANUAL DATE/TIME CHANGE FROM (B)(6) 2017 20:11 TO (B)(6) 2017 20:15. THERE WAS A PUMP REBOOTS ON (B)(6) 2017 AT 23:54 IN WHICH THE TIME/DATE WAS SET TO (B)(6) 2017 12:56 WHEN THE PUMP WAS TURNED BACK ON. EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. THE TOTAL DAILY DOSES ADDED UP CORRECTLY AND REFLECTED THE USERS PROGRAMMED BASAL RATES. THE PUMP PASSED A DELIVERY ACCURACY TEST AND WAS FOUND TO BE DELIVERING ACCURATELY AND WITHIN RANGE. UNRELATED TO THE ORIGINAL COMPLAINT, THE DISPLAY WAS DISCOLORED AND THE BATTERY COMPARTMENT WAS CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).

Additional Manufacturer Narrative · 1

THIS ISSUE IS NOT LIKELY TO CAUSE AN ADVERSE EVENT BECAUSE THE PUMP DISPLAYS THE VERIFY SCREEN AFTER IT IS REBOOTED AND THE TIME AND DATE MUST BE SET TO CONFIRM THE VERIFY SCREEN. THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2017, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT WAS TREATED IN THE EMERGENCY ROOM (ER) WITH A LOW BLOOD GLUCOSE (BG) OF 38 MG/DL WITH NO REPORTED SYMPTOMS ASSOCIATED WITH A TIME/DATE RESET ISSUE ON (B)(6) 2017. REPORTEDLY, THE PATIENT CONTINUED PUMP THERAPY AND WAS TREATED WITH INTRAVENOUS (IV) GLUCOSE AND RECEIVED FOOD AND DRINK AS TREATMENT (PEANUT BUTTER AND JELLY SANDWICH). DURING TROUBLESHOOTING WITH CUSTOMER TECHNICAL SUPPORT (CTS), IT WAS REVEALED THAT THE TIME AND DATE RESET TO DEFAULT WHEN THE BATTERY WAS REMOVED FROM THE PUMP FOR LESS THAN 24 HOURS. THE REPORTER STATED THAT WHEN CHANGING THE BATTERY THE USER DID NOT KNOW HOW TO CHANGE THE TIME FROM AM TO PM AND A FAMILY MEMBER CALLED THE AMBULANCE AT 4AM BECAUSE THE USER¿S BG WAS IN THE LOW 20'S. THE USER WAS THEN TAKEN TO THE ER BY AMBULANCE, RECEIVED THE STATED TREATMENT AND WAS RELEASED AFTER 4 HOURS. THIS COMPLAINT IS BEING REPORTED BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPOGLYCEMIA ASSOCIATED WITH A TIME/DATE RESET ISSUE, WITH USE ERROR AS A CONTRIBUTING FACTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
317939 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 57 YR Life Threatening| R