FDA Adverse Event Injury Summary report: N

RAYSTATION

MDR report key: 6529323 · Received April 28, 2017

Report

Report Number
3007774465-2017-00002
Event Type
Injury
Date Received
April 28, 2017
Date of Event
February 16, 2017
Report Date
April 4, 2025
Manufacturer
RAYSEARCH LABORATORIES AB (PUBL)
Product Code
MUJ
PMA / PMN Number
K160093
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

RAYSTATION WEDGE FACTORS ARE DEFINED AS WEDGE OUTPUT DIVIDED BY OUTPUT FOR THE SAME SIZE OPEN FIELD. THE WEDGE FACTOR DEFINITION IS CLEARLY STATED IN THE LABELING (REF. MANUAL 1.6.2) AND IS ALSO INCLUDED IN PHYSICS TRAINING.

Description of Event or Problem · 0

FOLLOW-UP OF REPORT RSL MDR: 3007774465-2017-00002 (B)(6) WEDGE FACTORS NORMALIZED TO 10X10 FIELD. A VETERINARY CLINIC USED A CONTRACT SERVICE TO HELP THEM WITH BEAM MODELING. THE USER CREATED THE BEAM MODEL USING WEDGE FACTORS CALCULATED AS WEDGE FIELD NORMALIZED TO THE 10X10 OPEN FIELD. HOWEVER, WEDGE FACTORS TO BE ENTERED IN THE RAYSTATION PHYSICS MODULE ARE DEFINED AS WEDGE OUTPUT DIVIDED BY OUTPUT FOR THE SAME SIZE OPEN FIELD. AS A RESULT, THE BEAM MODEL WAS INCORRECT FOR WEDGED FIELDS FOR FIELD SIZES OTHER THAN 10X10. THE RAYSTATION DEFINITION IS CLEARLY DESCRIBED IN THE RAYSTATION LABELING, BUT THE USER ASSUMED THAT THE DEFINITION WAS THE SAME AS FOR A SYSTEM HE/SHE HAD WORKED WITH PREVIOUSLY. WE DO NOT KNOW THE EXTENT OF BEAM MODEL VALIDATION OR PLAN QA DONE BY THE CLINIC. 7 ANIMAL PATIENTS WERE TREATED WITH WEDGES USING THE INCORRECT BEAM MODEL.

Description of Event or Problem · 0

FOLLOW-UP OF REPORT RSL MDR: 3007774465-2017-00002 (B)(6) WEDGE FACTORS NORMALIZED TO 10X10 FIELD. CLINICAL INCIDENT (MISTREATMENT). A VETERINARY CLINIC USED A CONTRACT SERVICE TO HELP THEM WITH BEAM MODELING. THE USER CREATED THE BEAM MODEL USING WEDGE FACTORS CALCULATED AS WEDGE FIELD NORMALIZED TO THE 10X10 OPEN FIELD. HOWEVER, WEDGE FACTORS TO BE ENTERED IN THE RAYSTATION PHYSICS MODULE ARE DEFINED AS WEDGE OUTPUT DIVIDED BY OUTPUT FOR THE SAME SIZE OPEN FIELD. AS A RESULT, THE BEAM MODEL WAS INCORRECT FOR WEDGED FIELDS FOR FIELD SIZES OTHER THAN 10X10. THE RAYSTATION DEFINITION IS CLEARLY DESCRIBED IN THE RAYSTATION LABELING, BUT THE USER ASSUMED THAT THE DEFINITION WAS THE SAME AS FOR A SYSTEM HE/SHE HAD WORKED WITH PREVIOUSLY. WE DO NOT KNOW THE EXTENT OF BEAM MODEL VALIDATION OR PLAN QA DONE BY THE CLINIC. 7 ANIMAL PATIENTS WERE TREATED WITH WEDGES USING THE INCORRECT BEAM MODEL.

Description of Event or Problem · 1

REPORT OF USE ERROR. A VETERINARY CLINIC USED A CONTRACT SERVICE TO HELP THEM WITH BEAM MODELING. THE USER CREATED THE BEAM MODEL USING WEDGE FACTORS CALCULATED AS WEDGE FIELD NORMALIZED TO THE 10X10 OPEN FIELD. HOWEVER, WEDGE FACTORS TO BE ENTERED IN THE RAYSTATION PHYSICS MODULE ARE DEFINED AS WEDGE OUTPUT DIVIDED BY OUTPUT FOR THE SAME SIZE OPEN FIELD. AS A RESULT, THE BEAM MODEL WAS INCORRECT FOR WEDGED FIELDS FOR FIELD SIZES OTHER THAN 10X10. THE RAYSTATION DEFINITION IS CLEARLY DESCRIBED IN THE RAYSTATION LABELING, BUT THE USER ASSUMED THAT THE DEFINITION WAS THE SAME AS FOR A SYSTEM HE/SHE HAD WORKED WITH PREVIOUSLY. WE DO NOT KNOW THE EXTENT OF BEAM MODEL VALIDATION OR PLAN QA DONE BY THE CLINIC. SEVEN ANIMAL PATIENTS WERE TREATED WITH WEDGES USING THE INCORRECT BEAM MODEL. WE CURRENTLY HAVE NO INFORMATION ON THE LEVEL OF DOSE DEVIATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311013 RAYSTATION RADIATION THERAPY TREATMENT PLANNING SYSTEM MUJ RAYSEARCH LABORATORIES AB (PUBL) RAYSTATION 5

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other