FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA

MDR report key: 6529083 · Received April 28, 2017

Report

Report Number
2954323-2017-03354
Event Type
Malfunction
Date Received
April 28, 2017
Date of Event
April 28, 2017
Report Date
January 30, 2018
Manufacturer
NBW
Product Code
NBW
PMA / PMN Number
K040814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

AN EXTENDED INVESTIGATION HAS BEEN CONDUCTED, WHICH INVOLVED A MANUFACTURING REVIEW (INCLUDING 5 YEARS OF DEVICE HISTORY RECORDS (DHRS)), PREVIOUS COMPLAINT AND CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATIONS CONDUCTED FOR BLANK SCREEN ISSUES, PROCESS FAILURE MODE EFFECTS ANALYSES (PFMEAS), DESIGN CONTROLS, AND DESIGN FAILURE MODE EFFECTS ANALYSES (DFMEAS), RISK MANAGEMENT REPORTS, RISK EVALUATIONS, AND LABEL COPY. THE INVESTIGATION DID NOT IDENTIFY ANY INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. IF THE PRODUCT IS RETURNED, AN INVESTIGATION WILL BE PERFORMED.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313098 PRECISION XTRA BLOOD GLUCOSE MONITORING SYSTEM NBW NBW 98814-65

Patients

Seq Age Sex Outcome Treatment
1